2 51

Cited 0 times in

FloWise Flow Diverter for Treatment of Unruptured Wide-Neck Intracranial Aneurysms: A Prospective, Multicenter, Single-Arm, Open-Label, Pivotal Study

Authors
 Hyun Jin Han  ;  Keun Young Park  ;  Byung Moon Kim  ;  Yong Bae Kim  ;  Hong-Jun Jeon  ;  Sang Hyun Suh  ;  Sangil Suh  ;  Jung-Jae Kim  ;  Jun-Hwee Kim  ;  Chang Ki Jang  ;  Dong Joon Kim  ;  Jung Keun Lee  ;  Sunghan Kim  ;  Kyu Seon Chung  ;  Solbi Kim  ;  Sun Yoon  ;  Min Jeoung Kim 
Citation
 KOREAN JOURNAL OF RADIOLOGY, Vol.26(7) : 716-726, 2025-07 
Journal Title
KOREAN JOURNAL OF RADIOLOGY
ISSN
 1229-6929 
Issue Date
2025-07
MeSH
Aged ; Cerebral Angiography ; Endovascular Procedures* / instrumentation ; Female ; Humans ; Intracranial Aneurysm* / diagnostic imaging ; Intracranial Aneurysm* / surgery ; Intracranial Aneurysm* / therapy ; Male ; Middle Aged ; Prospective Studies ; Stents* ; Treatment Outcome
Keywords
Flow diversion ; Intracranial aneurysm ; New device
Abstract
Objective: We evaluated the safety and efficacy of a new flow diverter in the treatment of unruptured wide-neck aneurysms.

Materials and methods: Patients who had unruptured intracranial aneurysms with a neck diameter ≥4 mm or dome-to-neck ratio <2 and who met other eligibility criteria were enrolled from seven institutions. The patients were treated with a new flow diverter (FloWise; Taewoong Medical, Seoul, Korea) and followed at 1, 3, 6, and 12 months after treatment. The primary efficacy endpoint was the complete occlusion of the target aneurysm at 12 months. Safety endpoints included ipsilateral stroke (National Institutes of Health Stroke Scale score ≥5), newly developed neurological deficit (modified Rankin Scale score ≥3), death at 30 days, and ≥50% parent artery stenosis or occlusion at the 12-month follow-up. The superiority of FloWise over the predefined reference value (56.7%) was investigated using a one-sided binomial test with a statistical significance of P < 0.025.

Results: Seventy patients (59.0 ± 9.9 years; male:female = 16:54) with 93 aneurysms (median diameter, 5.2 mm [range, 1.6-26.1 mm]; median neck diameter, 4.1 mm [range, 1.2-22.4 mm]) were included. Of these, 69 patients who completed the follow-up as planned constituted the efficacy analysis set. Complete aneurysm occlusion was achieved in 56 target aneurysms (81.2%; 95% confidence interval [CI], 69.9-89.6; P < 0.001). Procedural success and ≥50% aneurysm volume reduction were observed in 69 (100%) and 66 (95.7%; 95% CI, 87.8-99.1) target aneurysms, respectively. No ipsilateral strokes occurred in any of the 70 patients. Serious adverse events occurred in 6 patients; these events were causally related to the treatment in 4 (5.7%) cases and possibly related to the device in 2 (2.9%) cases.

Conclusion: FloWise flow diverter is safe and effective for the treatment of unruptured wide-neck aneurysms.
Files in This Item:
T202505445.pdf Download
DOI
10.3348/kjr.2024.1212
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiology (영상의학교실) > 1. Journal Papers
1. College of Medicine (의과대학) > Dept. of Neurosurgery (신경외과학교실) > 1. Journal Papers
Yonsei Authors
Kim, Dong Joon(김동준) ORCID logo https://orcid.org/0000-0002-7035-087X
Kim, Byung Moon(김병문) ORCID logo https://orcid.org/0000-0001-8593-6841
Kim, Yong Bae(김용배) ORCID logo https://orcid.org/0000-0003-2262-7157
Kim, Jung-Jae(김정재) ORCID logo https://orcid.org/0000-0002-4669-8577
Kim, Jun-Hwee(김준휘)
Park, Keun Young(박근영)
Suh, Sang Hyun(서상현) ORCID logo https://orcid.org/0000-0002-7098-4901
Jang, Chang Ki(장창기) ORCID logo https://orcid.org/0000-0001-8715-8844
Han, Hyun Jin(한현진) ORCID logo https://orcid.org/0000-0002-4111-4819
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/207198
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links