0 143

Cited 0 times in

Cited 19 times in

In vivo flexible energy harvesting on porcine heart via highly-piezoelectric PIN-PMN-PT single crystal

DC Field Value Language
dc.contributor.authorAn, Jaehun-
dc.contributor.authorPark, Hyewon-
dc.contributor.authorJung, Young Hoon-
dc.contributor.authorMin, Seongwook-
dc.contributor.authorKim, Dong Hyun-
dc.contributor.authorJoe, Daniel J.-
dc.contributor.authorLee, Sang-Goo-
dc.contributor.authorHyeon, Dong Yeol-
dc.contributor.authorJe, Yub-
dc.contributor.authorSeo, Hee-Seon-
dc.contributor.authorJeong, Uichang-
dc.contributor.authorHong, Seungbum-
dc.contributor.authorHwang, Geon-Tae-
dc.contributor.authorJoung, Boyoung-
dc.contributor.authorLee, Keon Jae-
dc.date.accessioned2025-02-03T08:45:40Z-
dc.date.available2025-02-03T08:45:40Z-
dc.date.created2025-02-14-
dc.date.issued2024-03-
dc.identifier.issn2211-2855-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/201832-
dc.description.abstractOne of the main challenges with implantable biomedical devices is the replacement surgeries of depleted batteries for elderly patients. This not only poses medical risks but also results in financial burdens. Consequently, there is a growing need for self-powered in vivo electronics that can convert the biomechanical movements of organs into usable electric energy. Notably, a previously reported in vivo flexible energy harvester generated relatively low current output from a living porcine heart, restricting its ability to operate self-powered electronic devices. In this study, a high current output implantable flexible energy harvester was demonstrated by adopting a highly-piezoelectric single crystal Pb(In1/2Nb1/2)O3-Pb(Mg1/3Nb2/3)O3-PbTiO3 (PIN-PMN-PT) film. The in vivo flexible energy harvester generated a short-circuit output current of 20 mu A (corresponding to a current density of 3.08 mu A/mm3) from the contraction and relaxation of a porcine heart. This recorded in vivo output current density was attributed to its outstanding piezoelectric charge coefficient of the single crystal PIN-PMNPT, optimized Ni stressor assisted exfoliation as well as the implementation of a metal-insulator-metal (MIM) structure. Moreover, the flexible energy harvester was also demonstrated as a self-powered cardiac sensor to monitor irregular heartbeats by observing heart rate variations due to drug administration in a porcine heart. Additionally, output performance evaluation was conducted with the chest closed to further assess the feasibility of practical in vivo harvesting. Finally, the single crystal harvester exhibited biocompatibility, showing no signs of cytotoxicity based on cell viability assessments and histological characterizations. These results underscore the potential of the flexible PIN-PMN-PT energy harvesting device as a sustainable power source for self-powered in vivo biomedical electronics.-
dc.description.statementOfResponsibilityrestriction-
dc.language영어-
dc.publisherELSEVIER-
dc.relation.isPartOfNANO ENERGY-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.titleIn vivo flexible energy harvesting on porcine heart via highly-piezoelectric PIN-PMN-PT single crystal-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Internal Medicine (내과학교실)-
dc.contributor.googleauthorAn, Jaehun-
dc.contributor.googleauthorPark, Hyewon-
dc.contributor.googleauthorJung, Young Hoon-
dc.contributor.googleauthorMin, Seongwook-
dc.contributor.googleauthorKim, Dong Hyun-
dc.contributor.googleauthorJoe, Daniel J.-
dc.contributor.googleauthorLee, Sang-Goo-
dc.contributor.googleauthorHyeon, Dong Yeol-
dc.contributor.googleauthorJe, Yub-
dc.contributor.googleauthorSeo, Hee-Seon-
dc.contributor.googleauthorJeong, Uichang-
dc.contributor.googleauthorHong, Seungbum-
dc.contributor.googleauthorHwang, Geon-Tae-
dc.contributor.googleauthorJoung, Boyoung-
dc.contributor.googleauthorLee, Keon Jae-
dc.identifier.doi10.1016/j.nanoen.2023.109227-
dc.subject.keywordImplantable self-powered device-
dc.subject.keywordIn vivo flexible energy harvesting-
dc.subject.keywordBiomechanical energy conversion-
dc.subject.keywordPiezoelectric single crystal-
dc.contributor.alternativeNameJoung, Bo Young-
dc.contributor.affiliatedAuthorPark, Hyewon-
dc.contributor.affiliatedAuthorJoung, Boyoung-
dc.identifier.scopusid2-s2.0-85181140046-
dc.identifier.wosid001165640100001-
dc.citation.volume121-
dc.identifier.bibliographicCitationNANO ENERGY, Vol.121, 2024-03-
dc.identifier.rimsid84746-
dc.type.rimsART-
dc.description.journalClass1-
dc.description.journalClass1-
dc.subject.keywordAuthorImplantable self-powered device-
dc.subject.keywordAuthorIn vivo flexible energy harvesting-
dc.subject.keywordAuthorBiomechanical energy conversion-
dc.subject.keywordAuthorPiezoelectric single crystal-
dc.subject.keywordPlusTHIN-FILM NANOGENERATOR-
dc.subject.keywordPlusPACEMAKER-
dc.subject.keywordPlusVITRO-
dc.type.docTypeArticle-
dc.description.isOpenAccessN-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalWebOfScienceCategoryChemistry, Physical-
dc.relation.journalWebOfScienceCategoryNanoscience & Nanotechnology-
dc.relation.journalWebOfScienceCategoryMaterials Science, Multidisciplinary-
dc.relation.journalWebOfScienceCategoryPhysics, Applied-
dc.relation.journalResearchAreaChemistry-
dc.relation.journalResearchAreaScience & Technology - Other Topics-
dc.relation.journalResearchAreaMaterials Science-
dc.relation.journalResearchAreaPhysics-
dc.identifier.articleno109227-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Internal Medicine (내과학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.