Cited 0 times in

Predictors of success in left bundle branch area pacing with stylet-driven pacing leads: a multicenter investigation

DC Field Value Language
dc.contributor.author김대훈-
dc.contributor.author김태훈-
dc.contributor.author박희남-
dc.contributor.author엄재선-
dc.contributor.author유희태-
dc.contributor.author이문형-
dc.contributor.author정보영-
dc.date.accessioned2024-12-06T01:54:45Z-
dc.date.available2024-12-06T01:54:45Z-
dc.date.issued2024-09-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/200634-
dc.description.abstractPurpose: Although left bundle branch area pacing (LBBAP) is an emerging conduction system pacing modality, it is unclear which parameters predict procedural success and how many implant attempts are acceptable. This study aimed to assess predictors of successful LBBAP, left bundle branch (LBB) capture, and factors associated with the number of LBBAP implant attempts. Methods: This retrospective observational multicenter study was conducted in Korea. LBBAP was attempted in 119 patients; 89.3% of patients had bradyarrhythmia (atrioventricular block 82.4%), and 10.7% of patients had heart failure (cardiac resynchronization therapy) indication. Procedural success and electrophysiological and echocardiographic parameters were evaluated. Results: The acute success rate of lead implantation in LBBAP was 95.8% (114 of 119 patients) and that of LBB capture was 82.4% (98 of 119 patients). Fewer implant attempts were associated with LBBAP success (three or fewer vs. over three times, p = 0.014) and LBB capture (three or fewer vs. over three times, p = 0.010). In the multivariate linear regression, the patients with intraventricular conduction delay (IVCD) required a greater number of attempts than those without IVCD [estimates = 2.33 (0.35–4.31), p = 0.02], and the larger the right atrial (RA) size, the more the attempts required for LBBAP lead implantation [estimates = 2.08 (1.20–2.97), p < 0.001]. Conclusion: An increase in the number of implant attempts was associated with LBBAP procedural failure and LBB capture failure. The electrocardiographic parameter IVCD and the echocardiographic parameter RA size may predict the procedural complexity and the number of lead implant attempts for LBBAP.-
dc.description.statementOfResponsibilityopen-
dc.languageEnglish-
dc.publisherFrontiers Media S.A.-
dc.relation.isPartOfFRONTIERS IN CARDIOVASCULAR MEDICINE-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.titlePredictors of success in left bundle branch area pacing with stylet-driven pacing leads: a multicenter investigation-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Internal Medicine (내과학교실)-
dc.contributor.googleauthorGa-In Yu-
dc.contributor.googleauthorTae-Hoon Kim-
dc.contributor.googleauthorJung-Myung Lee-
dc.contributor.googleauthorDaehoon Kim-
dc.contributor.googleauthorHee Tae Yu-
dc.contributor.googleauthorJae-Sun Uhm-
dc.contributor.googleauthorBoyoung Joung-
dc.contributor.googleauthorHui-Nam Pak-
dc.contributor.googleauthorMoon-Hyoung Lee-
dc.identifier.doi10.3389/fcvm.2024.1449859-
dc.contributor.localIdA00373-
dc.contributor.localIdA01085-
dc.contributor.localIdA01776-
dc.contributor.localIdA02337-
dc.contributor.localIdA02535-
dc.contributor.localIdA02766-
dc.contributor.localIdA03609-
dc.relation.journalcodeJ04002-
dc.identifier.eissn2297-055X-
dc.identifier.pmid39376621-
dc.subject.keywordconduction system pacing-
dc.subject.keywordinterventricular conduction delay-
dc.subject.keywordleft bundle branch area pacing-
dc.subject.keywordright atrial diameter-
dc.subject.keywordstylet-driven pacing leads-
dc.contributor.alternativeNameKim, Dae Hoon-
dc.contributor.affiliatedAuthor김대훈-
dc.contributor.affiliatedAuthor김태훈-
dc.contributor.affiliatedAuthor박희남-
dc.contributor.affiliatedAuthor엄재선-
dc.contributor.affiliatedAuthor유희태-
dc.contributor.affiliatedAuthor이문형-
dc.contributor.affiliatedAuthor정보영-
dc.citation.volume11-
dc.citation.startPage1449859-
dc.identifier.bibliographicCitationFRONTIERS IN CARDIOVASCULAR MEDICINE, Vol.11 : 1449859, 2024-09-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Internal Medicine (내과학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.