0 243

Cited 0 times in

Cited 0 times in

Automation of 3D digital rolling circle amplification using a 3D-printed liquid handler

DC Field Value Language
dc.contributor.author정해림-
dc.contributor.author조현수-
dc.date.accessioned2024-08-19T00:04:48Z-
dc.date.available2024-08-19T00:04:48Z-
dc.date.issued2024-10-
dc.identifier.issn0956-5663-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/200207-
dc.description.abstractAutomation of liquid handling is indispensable to improve throughput and reproducibility in biochemical assays. However, the incorporation of automated systems into laboratory workflows is often hindered by the high cost and complexity associated with building robotic liquid handlers. Here, we report a 3D-printed liquid handler based on a fluidic manifold, thereby obviating the need for complex robotic mechanisms. The fluidic manifold, termed a dispensing and aspirating (DA) device, comprises parallelized multi-pipette structures connected by distribution and aspiration channels, enabling the precise supply and removal of reagents, respectively. Leveraging the versatility of 3D printing, the DA device can be custom-designed and printed to fit specific applications. As a proof-of-principle, we engineered a 3D-printed liquid handler dedicated for 3D digital rolling circle amplification (4DRCA), an advanced biochemical assay involving multiple sample preparation steps such as antibody incubation, cell fixation, nucleic acid amplification, probe hybridization, and extensive washing. We demonstrate the efficacy of the 3D-printed liquid handler to automate the preparation of clinical samples for the simultaneous, in situ analysis of oncogenic protein and transcript markers in B-cell acute lymphoblastic leukemia cells using 4DRCA. This approach provides an effective and accessible solution for liquid handling automation, offering high throughput and reproducibility in biochemical assays.-
dc.description.statementOfResponsibilityrestriction-
dc.languageEnglish-
dc.publisherElsevier Advanced Technology-
dc.relation.isPartOfBIOSENSORS & BIOELECTRONICS-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.subject.MESHAutomation-
dc.subject.MESHBiosensing Techniques* / instrumentation-
dc.subject.MESHBiosensing Techniques* / methods-
dc.subject.MESHEquipment Design-
dc.subject.MESHHumans-
dc.subject.MESHNucleic Acid Amplification Techniques* / instrumentation-
dc.subject.MESHNucleic Acid Amplification Techniques* / methods-
dc.subject.MESHPrinting, Three-Dimensional*-
dc.titleAutomation of 3D digital rolling circle amplification using a 3D-printed liquid handler-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Internal Medicine (내과학교실)-
dc.contributor.googleauthorSuyeon Shin-
dc.contributor.googleauthorHyo Geun Yun-
dc.contributor.googleauthorHaerim Chung-
dc.contributor.googleauthorHyunsoo Cho-
dc.contributor.googleauthorSungyoung Choi-
dc.identifier.doi10.1016/j.bios.2024.116503-
dc.contributor.localIdA04674-
dc.contributor.localIdA03929-
dc.relation.journalcodeJ00330-
dc.identifier.eissn1873-4235-
dc.identifier.pmid38905856-
dc.identifier.urlhttps://www.sciencedirect.com/science/article/pii/S0956566324005086-
dc.subject.keyword3D digital rolling circle amplification-
dc.subject.keyword3D printing-
dc.subject.keywordLeukemia diagnosis-
dc.subject.keywordLiquid handler-
dc.contributor.alternativeNameChung, Hae Rim-
dc.contributor.affiliatedAuthor정해림-
dc.contributor.affiliatedAuthor조현수-
dc.citation.volume261-
dc.citation.startPage116503-
dc.identifier.bibliographicCitationBIOSENSORS & BIOELECTRONICS, Vol.261 : 116503, 2024-10-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Internal Medicine (내과학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.