18 61

Cited 0 times in

Development of shielding evaluation and management program for O-ring type linear accelerators

DC Field Value Language
dc.contributor.author김동욱-
dc.contributor.author김지훈-
dc.contributor.author김진성-
dc.date.accessioned2024-06-14T03:14:38Z-
dc.date.available2024-06-14T03:14:38Z-
dc.date.issued2024-05-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/199815-
dc.description.abstractThe shielding parameters can vary depending on the geometrical structure of the linear accelerators (LINAC), treatment techniques, and beam energies. Recently, the introduction of O-ring type linear accelerators is increasing. The objective of this study is to evaluate the shielding parameters of new type of linac using a dedicated program developed by us named ORSE (O-ring type Radiation therapy equipment Shielding Evaluation). The shielding evaluation was conducted for a total of four treatment rooms including Elekta Unity, Varian Halcyon, and Accuray Tomotherapy. The developed program possesses the capability to calculate transmitted dose, maximum treatable patient capacity, and shielding wall thickness based on patient data. The doses were measured for five days using glass dosimeters to compare with the results of program. The IMRT factors and use factors obtained from patient data showed differences of up to 65.0% and 33.8%, respectively, compared to safety management report. The shielding evaluation conducted in each treatment room showed that the transmitted dose at every location was below 1% of the dose limit. The results of program and measurements showed a maximum difference of 0.003 mSv/week in transmitted dose. The ORSE program allows for the shielding evaluation results to the clinical environment of each institution based on patient data. © The Author(s) 2024.-
dc.description.statementOfResponsibilityopen-
dc.languageEnglish-
dc.publisherNature Publishing Group-
dc.relation.isPartOfSCIENTIFIC REPORTS-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.subject.MESHHumans-
dc.subject.MESHParticle Accelerators* / instrumentation-
dc.subject.MESHRadiation Dosage-
dc.subject.MESHRadiation Protection* / instrumentation-
dc.subject.MESHRadiation Protection* / methods-
dc.subject.MESHRadiotherapy, Intensity-Modulated / methods-
dc.titleDevelopment of shielding evaluation and management program for O-ring type linear accelerators-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Radiation Oncology (방사선종양학교실)-
dc.contributor.googleauthorDong Hyeok Choi-
dc.contributor.googleauthorSo Hyun Ahn-
dc.contributor.googleauthorDong Wook Kim-
dc.contributor.googleauthorSang Hyoun Choi-
dc.contributor.googleauthorWoo Sang Ahn-
dc.contributor.googleauthorJihun Kim-
dc.contributor.googleauthorJin Sung Kim-
dc.identifier.doi10.1038/s41598-024-60362-6-
dc.contributor.localIdA05710-
dc.contributor.localIdA05823-
dc.contributor.localIdA04548-
dc.relation.journalcodeJ02646-
dc.identifier.eissn2045-2322-
dc.identifier.pmid38729975-
dc.subject.keywordDICOM-RT-
dc.subject.keywordHalcyon-
dc.subject.keywordShielding evaluation-
dc.subject.keywordTomotherapy-
dc.subject.keywordUnity-
dc.contributor.alternativeNameKim, Dong Wook-
dc.contributor.affiliatedAuthor김동욱-
dc.contributor.affiliatedAuthor김지훈-
dc.contributor.affiliatedAuthor김진성-
dc.citation.volume14-
dc.citation.number1-
dc.citation.startPage10719-
dc.identifier.bibliographicCitationSCIENTIFIC REPORTS, Vol.14(1) : 10719, 2024-05-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiation Oncology (방사선종양학교실) > 1. Journal Papers

qrcode