119 273

Cited 0 times in

Cited 0 times in

Dosimetric verification of annual quality assurance for a linear accelerator using a transmission type detector

DC Field Value Language
dc.contributor.authorChoi, Dong Hyeok-
dc.contributor.authorKim, Jin Sung-
dc.contributor.authorLee, Rena-
dc.contributor.authorAhn, So Hyun-
dc.contributor.authorAhn, Woo Sang-
dc.date.accessioned2024-03-22T05:52:01Z-
dc.date.available2024-03-22T05:52:01Z-
dc.date.created2024-04-05-
dc.date.issued2023-10-
dc.identifier.issn2045-2322-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/198263-
dc.description.abstractThe purpose of our study is to establish an efficient quality assurance (QA) procedure using a transmission-type detector (IBA, Stealth chamber), a reference signal detector, as a field chamber. Relative dosimetry items, including monitor unit linearity, output constancy based on dose rate and field size, and output factor were measured and compared with results obtained from the Farmer-type chamber (IBA, Wellhofer, FC65-G). Moreover, output for each field size was measured to assess its applicability to small fields. Results using the Stealth chamber were in good agreement with the FC65-G within 1.0%, except for output constancy according to gantry angle, which had a 1.1% error rate for the Stealth chamber and 2.7% for the FC65-G. Differences of up to - 6.26% output factor were observed for the Stealth chamber and up to - 0.56% for the CC-13 ionization chamber (IBA) in the 3 x 3 cm(2 )field. Our study confirmed the possibility of using Stealth chambers for relative dosimetry measurement in QA.-
dc.description.statementOfResponsibilityopen-
dc.formatapplication/pdf-
dc.languageEnglish-
dc.publisherNature Publishing Group-
dc.relation.isPartOfSCIENTIFIC REPORTS-
dc.relation.isPartOfSCIENTIFIC REPORTS-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.titleDosimetric verification of annual quality assurance for a linear accelerator using a transmission type detector-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Radiation Oncology (방사선종양학교실)-
dc.contributor.googleauthorChoi, Dong Hyeok-
dc.contributor.googleauthorKim, Jin Sung-
dc.contributor.googleauthorLee, Rena-
dc.contributor.googleauthorAhn, So Hyun-
dc.contributor.googleauthorAhn, Woo Sang-
dc.identifier.doi10.1038/s41598-023-45114-2-
dc.relation.journalcodeJ02646-
dc.identifier.eissn2045-2322-
dc.identifier.pmid37865666-
dc.contributor.alternativeNameKim, Jinsung-
dc.contributor.affiliatedAuthorKim, Jin Sung-
dc.identifier.scopusid2-s2.0-85174616880-
dc.identifier.wosid001094273200036-
dc.citation.volume13-
dc.citation.number1-
dc.identifier.bibliographicCitationSCIENTIFIC REPORTS, Vol.13(1), 2023-10-
dc.identifier.rimsid82876-
dc.type.rimsART-
dc.description.journalClass1-
dc.description.journalClass1-
dc.subject.keywordPlusGANTRY ANGLE-
dc.subject.keywordPlusBEAMS-
dc.type.docTypeArticle-
dc.description.isOpenAccessY-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalWebOfScienceCategoryMultidisciplinary Sciences-
dc.relation.journalResearchAreaScience & Technology - Other Topics-
dc.identifier.articleno17994-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiation Oncology (방사선종양학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.