Cited 8 times in
Accuracy of dental implant placement with computer-guided surgery: a retrospective cohort study
DC Field | Value | Language |
---|---|---|
dc.contributor.author | 구정귀 | - |
dc.date.accessioned | 2023-06-02T00:39:01Z | - |
dc.date.available | 2023-06-02T00:39:01Z | - |
dc.date.issued | 2022-01 | - |
dc.identifier.uri | https://ir.ymlib.yonsei.ac.kr/handle/22282913/194331 | - |
dc.description.abstract | Background: The aim of this study was to assess the accuracy of virtual planning of computer-guided surgery based on the actual outcomes of clinical dental implant placement. Methods: This retrospective study enrolled patients among whom implant treatment was planned using computer-guided surgery with cone beam computed tomography (CBCT). The patients who received implant according to the guide with the flapless and flapped approach were classified as group 1 and 2, respectively, and the others who could not be placed according to the guide were allocated to the drop-out group. The accuracy of implant placement was evaluated with the superimposition of CBCT. Results: We analyzed differences in the deviated distance of the entrance point and deviated angulation of the insertion of implant fixtures. With regard to the surgical approach, group 2 exhibited greater accuracy compared to group 1 in deviation distance (2.22 ± 0.88 and 3.18 ± 0.89 mm, respectively, P < 0.001) and angulation (4.27 ± 2.30 and 6.82 ± 2.71°, respectively, P = 0.001). The limitations of guided surgery were discussed while considering the findings from the drop-out group. Conclusions: Computer-guided surgery demonstrates greater accuracy in implant placement with the flapless approach. Further research should be conducted to enhance the availability of guides for cases with unfavorable residual bone conditions. | - |
dc.description.statementOfResponsibility | open | - |
dc.language | English | - |
dc.publisher | BioMed Central | - |
dc.relation.isPartOf | BMC ORAL HEALTH | - |
dc.rights | CC BY-NC-ND 2.0 KR | - |
dc.subject.MESH | Computer-Aided Design | - |
dc.subject.MESH | Computers | - |
dc.subject.MESH | Cone-Beam Computed Tomography | - |
dc.subject.MESH | Dental Implantation, Endosseous / methods | - |
dc.subject.MESH | Dental Implants* | - |
dc.subject.MESH | Humans | - |
dc.subject.MESH | Imaging, Three-Dimensional | - |
dc.subject.MESH | Patient Care Planning | - |
dc.subject.MESH | Retrospective Studies | - |
dc.subject.MESH | Surgery, Computer-Assisted* / methods | - |
dc.title | Accuracy of dental implant placement with computer-guided surgery: a retrospective cohort study | - |
dc.type | Article | - |
dc.contributor.college | College of Dentistry (치과대학) | - |
dc.contributor.department | Dept. of Oral and Maxillofacial Surgery (구강악안면외과학교실) | - |
dc.contributor.googleauthor | Jeong-Kui Ku | - |
dc.contributor.googleauthor | Junggon Lee | - |
dc.contributor.googleauthor | Hyo-Jung Lee | - |
dc.contributor.googleauthor | Pil-Young Yun | - |
dc.contributor.googleauthor | Young-Kyun Kim | - |
dc.identifier.doi | 10.1186/s12903-022-02046-z | - |
dc.contributor.localId | A06085 | - |
dc.relation.journalcode | J00371 | - |
dc.identifier.eissn | 1472-6831 | - |
dc.identifier.pmid | 35034613 | - |
dc.subject.keyword | Accuracy | - |
dc.subject.keyword | Computer-guided surgery | - |
dc.subject.keyword | Dental implant | - |
dc.subject.keyword | Implant-guided surgery | - |
dc.subject.keyword | Virtual plan | - |
dc.contributor.alternativeName | Ku, Jeong-Kui | - |
dc.contributor.affiliatedAuthor | 구정귀 | - |
dc.citation.volume | 22 | - |
dc.citation.number | 1 | - |
dc.citation.startPage | 8 | - |
dc.identifier.bibliographicCitation | BMC ORAL HEALTH, Vol.22(1) : 8, 2022-01 | - |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.