Performance Evaluation of the DxC 700 AU Chemistry Analyzer in Hemoglobin A1c Measurement

Authors
 Yu Jeong Choi  ;   Hyein Kang  ;   Chan-Ik Cho  ;   John Hoon Rim  ;   Sang-Guk Lee  ;   Jong-Baeck Lim  
Citation
 ANNALS OF LABORATORY MEDICINE, Vol.43(2) : 167-173, 2023-03 
Journal Title
ANNALS OF LABORATORY MEDICINE
ISSN
 2234-3806 
Issue Date
2023-03
MeSH
Chromatography, High Pressure Liquid ; Glycated Hemoglobin / analysis ; Hematologic Tests* ; Humans ; Immunoassay / methods
Keywords
Fetal hemoglobin ; Hemoglobin A1c ; Variant hemoglobin
Abstract
Background: Accurate measurement of glycated hemoglobin (HbA1c) is crucial for a diabetes diagnosis and subsequent patient management. The detection method and presence of variant Hb can interfere with HbA1c measurements. We evaluated the HbA1c-measuring performance of the DxC 700 AU (Beckman Coulter, Brea, CA, USA) immunoassay-based device in comparison with another immunoassay device and the reference method.

Methods: A total of 120 normal and 14 variant Hb samples were analyzed using the Cobas c 513 (Roche Diagnostics, Mannheim, Germany) and DxC 700 AU analyzers. Variant Hb samples were also analyzed using the reference method, along with 20 normal samples. The accuracy, precision, linearity, and carryover were determined.

Results: DxC 700 AU results strongly correlated with those of Cobas c 513 and exhibited accuracy in comparison with the reference method. The within-run, between-run, between-day, and total imprecision (%CV) values for the low- and high-concentration control materials were below 2%. The results of DxC 700 AU were linear over a wide HbA1c range (3.39%-18.30%). Although DxC 700 AU performed well in the presence of variant Hb, the HbA1c concentration was underestimated in the presence of fetal Hb. The possibility of interference from a high HbH proportion could not be ruled out.

Conclusions: The overall analytical performance of DxC 700 AU was acceptable. The device is accurate, precise, and linear over a wide HbA1c concentration range. Although DxC 700 AU results highly correlated with those of Cobas c 513, caution should be exercised in cases of high HbF and HbH concentrations.
Files in This Item:
T202301054.pdf Download
DOI
10.3343/alm.2023.43.2.167
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Laboratory Medicine (진단검사의학교실) > 1. Journal Papers
Yonsei Authors
Kang, Hyein(강혜인)
Lee, Sang-Guk(이상국) ORCID logo https://orcid.org/0000-0003-3862-3660
Rim, John Hoon(임정훈) ORCID logo https://orcid.org/0000-0001-6825-8479
Lim, Jong Baeck(임종백) ORCID logo https://orcid.org/0000-0003-0419-0422
Choi, Yu Jeong(최유정)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/193619
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links