231 519

Cited 0 times in

Cited 2 times in

Comprehensive clinical evaluation of TomoEQA for patient-specific pre-treatment quality assurance in helical tomotherapy

DC Field Value Language
dc.contributor.authorKim, Changhwan-
dc.contributor.authorHan, Mincheol-
dc.contributor.authorLee, Young Kyu-
dc.contributor.authorShin, Han Back-
dc.contributor.authorKim, HoJin-
dc.contributor.authorKim, Jin sung-
dc.date.accessioned2022-12-22T05:13:27Z-
dc.date.available2022-12-22T05:13:27Z-
dc.date.created2023-01-16-
dc.date.issued2022-11-
dc.identifier.issn1748-717X-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/192334-
dc.description.abstractBackground Based on a previous study on the feasibility of TomoEQA, an exit detector-based patient-specific pre-treatment quality assurance (QA) method for helical tomotherapy, an in-depth clinical evaluation was conducted. Methods Data of one hundred patients were analyzed to evaluate the clinical usefulness of TomoEQA for patient-specific pre-treatment QA in comparison with the conventional phantom-based method. Additional investigations were also performed under unusual measurement conditions to validate the off-axis region. In addition to the clinical evaluation of TomoEQA, a statistical analysis was conducted to determine the plan parameters that affect the pass/failure results of pre-treatment QA. Results The average and standard deviations of the gamma passing rate and point dose error for TomoEQA were comparable to those of the conventional QA method. For TomoEQA, the average values of the gamma passing rate and point dose error were 96.32% (standard deviation (1 sigma) = 3.94; 95% confidence interval (CI), 95.55 to 97.09) and - 1.12% (standard deviation (1 sigma) = 1.04; CI, - 1.32 to - 0.92), respectively. For the conventional QA method, the average values of the gamma passing rate and point dose error were 95.95% (standard deviation (1 sigma) = 4.35; 95% confidence interval (CI), 95.10 to 96.80) and - 1.20% (standard deviation (1 sigma) = 1.61; CI, - 1.52 to - 0.88), respectively. Further experiments on the off-axis region demonstrated that TomoEQA can provide accurate results for 3D dose analysis, which is inherently difficult in the conventional QA method. Through a statistical analysis based on the results of TomoEQA, it was validated that the total fraction (Total Fx), monitor units, beam-on-time, leaf-of-time below 100 ms, and planning target volume diameter were statistically significant for the pass/failure of the pre-treatment QA results. Conclusions TomoEQA is a clinically beneficial alternative to the conventional phantom-based QA method.-
dc.description.statementOfResponsibilityopen-
dc.languageEnglish-
dc.publisherBioMed Central-
dc.relation.isPartOfRadiation Oncology-
dc.relation.isPartOfRADIATION ONCOLOGY-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.titleComprehensive clinical evaluation of TomoEQA for patient-specific pre-treatment quality assurance in helical tomotherapy-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Radiation Oncology (방사선종양학교실)-
dc.contributor.googleauthorKim, Changhwan-
dc.contributor.googleauthorHan, Mincheol-
dc.contributor.googleauthorLee, Young Kyu-
dc.contributor.googleauthorShin, Han Back-
dc.contributor.googleauthorKim, HoJin-
dc.contributor.googleauthorKim, Jin sung-
dc.identifier.doi10.1186/s13014-022-02151-x-
dc.relation.journalcodeJ02591-
dc.identifier.eissn1748-717X-
dc.identifier.pmid36344983-
dc.subject.keywordHelical tomotherapy-
dc.subject.keywordPatient-specific pre-treatment quality assurance-
dc.subject.keywordExit detector-
dc.contributor.alternativeNameKim, Jinsung-
dc.contributor.affiliatedAuthorKim, Changhwan-
dc.contributor.affiliatedAuthorHan, Mincheol-
dc.contributor.affiliatedAuthorShin, Han Back-
dc.contributor.affiliatedAuthorKim, HoJin-
dc.contributor.affiliatedAuthorKim, Jin sung-
dc.identifier.scopusid2-s2.0-85141496608-
dc.identifier.wosid000879770100002-
dc.citation.volume17-
dc.citation.number1-
dc.identifier.bibliographicCitationRadiation Oncology, Vol.17(1), 2022-11-
dc.identifier.rimsid76134-
dc.type.rimsART-
dc.description.journalClass1-
dc.description.journalClass1-
dc.subject.keywordAuthorHelical tomotherapy-
dc.subject.keywordAuthorPatient-specific pre-treatment quality assurance-
dc.subject.keywordAuthorExit detector-
dc.subject.keywordPlusTREATMENT PLANNING PARAMETERS-
dc.subject.keywordPlusDELTA(4) PHANTOM-
dc.subject.keywordPlusQA SYSTEM-
dc.subject.keywordPlusIMRT-
dc.subject.keywordPlusVERIFICATION-
dc.subject.keywordPlusDELIVERY-
dc.subject.keywordPlusDOSIMETRY-
dc.subject.keywordPlusDOCUMENT-
dc.type.docTypeArticle-
dc.description.isOpenAccessY-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalWebOfScienceCategoryOncology-
dc.relation.journalWebOfScienceCategoryRadiology, Nuclear Medicine & Medical Imaging-
dc.relation.journalResearchAreaOncology-
dc.relation.journalResearchAreaRadiology, Nuclear Medicine & Medical Imaging-
dc.identifier.articleno177-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiation Oncology (방사선종양학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.