126 306

Cited 7 times in

Improving Left Atrial Appendage Occlusion Device Size Determination by Three-Dimensional Printing-Based Preprocedural Simulation

Authors
 William D Kim  ;  Iksung Cho  ;  Young Doo Kim  ;  Min Jae Cha  ;  Sang-Wook Kim  ;  Young Choi  ;  Seung Yong Shin 
Citation
 FRONTIERS IN CARDIOVASCULAR MEDICINE, Vol.9 : 830062, 2022-02 
Journal Title
FRONTIERS IN CARDIOVASCULAR MEDICINE
Issue Date
2022-02
Keywords
atrial fibrillation ; cardiac computed tomography ; left atrial appendage occlusion ; simulation ; three-dimensional printing
Abstract
Background: The two-dimensional (2D)-based left atrial appendage (LAA) occluder (LAAO) size determination by using transesophageal echocardiography (TEE) is limited by the structural complexity and wide anatomical variation of the LAA.

Objective: This study aimed to assess the accuracy of the LAAO size determination by implantation simulation by using a three-dimensional (3D)-printed model compared with the conventional method based on TEE.

Methods: We retrospectively reviewed patients with anatomically and physiologically properly implanted the Amplatzer Cardiac Plug and Amulet LAAO devices between January 2014 and December 2018 by using the final size of the implanted devices as a standard for size prediction accuracy. The use of 3D-printed model simulations in device sizing was compared with the conventional TEE-based method.

Results: A total of 28 cases with the percutaneous LAA occlusion were reviewed. There was a minimal difference [-0.11 mm; 95% CI (-0.93, 0.72 mm); P = 0.359] between CT-based reconstructed 3D images and 3D-printed left atrium (LA) models. Device size prediction based on TEE measurements showed poor agreement (32.1%), with a mean difference of 2.3 ± 3.2 mm [95% CI (-4.4, 9.0)]. The LAAO sizing by implantation simulation with 3D-printed models showed excellent correlation with the actually implanted LAAO size (r = 0.927; bias = 0.7 ± 2.5). The agreement between the 3D-printed and the implanted size was 67.9%, with a mean difference of 0.6 mm [95% CI (-1.9, 3.2)].

Conclusion: The use of 3D-printed LA models in the LAAO size determination showed improvement in comparison with conventional 2D TEE method.
Files in This Item:
T202200923.pdf Download
DOI
10.3389/fcvm.2022.830062
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Internal Medicine (내과학교실) > 1. Journal Papers
Yonsei Authors
Cho, Ik Sung(조익성)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/188347