393 678

Cited 5 times in

Cited 3 times in

Performance Evaluation of the KRYPTOR Compact PLUS Analyzer-Based B.R.A.H.M.S. CgA II KRYPTOR Assay for Chromogranin A Measurement

DC Field Value Language
dc.contributor.authorChoi, Yu Jeong-
dc.contributor.authorRoh, Juhye-
dc.contributor.authorKim, Sin Young-
dc.contributor.authorLee, Kyung A-
dc.contributor.authorPark, Youn Hee-
dc.date.accessioned2022-02-23T01:19:34Z-
dc.date.available2022-02-23T01:19:34Z-
dc.date.created2022-03-04-
dc.date.issued2021-12-
dc.identifier.issn2075-4418-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/187636-
dc.description.abstractNumerous immunoassays have been developed to measure the levels of chromogranin A (CgA), a useful biomarker for diagnosing and monitoring generally heterogeneous neuroendocrine tumors (NETs). Here, we evaluated the imprecision and linearity of three such assays: KRYPTOR (ThermoFisher Scientific), NEOLISA (EuroDiagnostica), and CgA-RIA (CisBio), using 123 samples for each assay. The correlation coefficients between the assays were 0.932 (CgA-RIA versus NEOLISA), 0.956 (KRYPTOR versus CgA-RIA), and 0.873 (NEOLISA versus KRYPTOR). KRYPTOR showed good precision, with percent coefficients of variation less than 5% for low and high concentration quality controls. Linearity was maintained over a wide concentration range. Comparison of CgA levels from three disease entities (NETs, non-NET pancreatic tumors, and prostate cancer) and healthy controls showed that patients with NETs had significantly higher CgA levels (n = 57, mean: 1.82 +/- 0.43 log ng/mL) than healthy individuals (n = 20, mean: 1.51 +/- 0.23 log ng/mL; p = 0.018). No other significant differences between groups were observed. All three immunoassays showed strong correlations in measured CgA levels. Because KRYPTOR operation uses a fully automated random-access system and requires shorter incubation times and smaller sample volumes, the KRYPTOR assay may improve laboratory workflow while maintaining satisfactory analytical performance.-
dc.description.statementOfResponsibilityopen-
dc.formatapplication/pdf-
dc.languageEnglish-
dc.publisherMDPI AG-
dc.relation.isPartOfDiagnostics-
dc.relation.isPartOfDIAGNOSTICS-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.titlePerformance Evaluation of the KRYPTOR Compact PLUS Analyzer-Based B.R.A.H.M.S. CgA II KRYPTOR Assay for Chromogranin A Measurement-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Laboratory Medicine (진단검사의학교실)-
dc.contributor.googleauthorChoi, Yu Jeong-
dc.contributor.googleauthorRoh, Juhye-
dc.contributor.googleauthorKim, Sin Young-
dc.contributor.googleauthorLee, Kyung A-
dc.contributor.googleauthorPark, Youn Hee-
dc.identifier.doi10.3390/diagnostics11122400-
dc.relation.journalcodeJ03798-
dc.identifier.eissn2075-4418-
dc.subject.keywordchromogranin A-
dc.subject.keywordneuroendocrine tumor-
dc.subject.keywordbiomarker-
dc.contributor.alternativeNameKim, Sin Young-
dc.contributor.affiliatedAuthorChoi, Yu Jeong-
dc.contributor.affiliatedAuthorKim, Sin Young-
dc.contributor.affiliatedAuthorLee, Kyung A-
dc.contributor.affiliatedAuthorPark, Youn Hee-
dc.identifier.scopusid2-s2.0-85121497797-
dc.identifier.wosid000736619400001-
dc.citation.volume11-
dc.citation.number12-
dc.identifier.bibliographicCitationDiagnostics, Vol.11(12), 2021-12-
dc.identifier.rimsid72785-
dc.type.rimsART-
dc.description.journalClass1-
dc.description.journalClass1-
dc.subject.keywordAuthorchromogranin A-
dc.subject.keywordAuthorneuroendocrine tumor-
dc.subject.keywordAuthorbiomarker-
dc.subject.keywordPlusB SECRETOGRANIN-I-
dc.subject.keywordPlusSERUM-LEVELS-
dc.subject.keywordPlusBIOMARKER-
dc.subject.keywordPlusPATHWAY-
dc.subject.keywordPlusSYSTEM-
dc.type.docTypeArticle-
dc.description.isOpenAccessY-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalWebOfScienceCategoryMedicine, General & Internal-
dc.relation.journalResearchAreaGeneral & Internal Medicine-
dc.identifier.articleno2400-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Laboratory Medicine (진단검사의학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.