0 494

Cited 2 times in

Assessment of dosimetric leaf gap correction factor in Mobius3D commissioning affected by couch top

DC Field Value Language
dc.contributor.author김동욱-
dc.contributor.author김지훈-
dc.contributor.author김진성-
dc.contributor.author박광우-
dc.contributor.author장경환-
dc.contributor.author한민철-
dc.contributor.author한수철-
dc.contributor.author홍채선-
dc.date.accessioned2020-12-01T17:37:13Z-
dc.date.available2020-12-01T17:37:13Z-
dc.date.issued2020-09-
dc.identifier.issn2662-4729-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/180353-
dc.description.abstractThis study assesses the dosimetric leaf gap (DLG) correction factor in Mobius3D commissioning affected by a couch top platform and calculates the optimal DLG value according to the point dose difference function. DLG optimizations were performed for 3 LINAC machines and a total of 30 patient volumetric modulated arc therapy plans (i.e., 10 plans per each LINAC). Point dose calculations were performed using an automatic dose calculation system in Mobius3D as well as Mobis3D calculation using a Mobius Verification Phantom (MVP)-based quality assurance plan with a carbon fiber couch top. Subsequently, the results were compared with measurement data. The averaged point dose measured for the MVP with a couch top decreased by approximately 2% relative to that without the couch top. The average of the optimal DLG factors increased by 1.153 mm due to the couch top effect for a dose decrease of 2% at the measured point. In the procedure of Mobius beam commissioning, users should adjust the DLG correction factor using a specific phantom (including MVP) with a couch top structure. If the DLG optimization were performed by using MVP automatic dose calculation system, the factor should be increased by approximately 1.2 mm per 2% dose difference considering user's couch top effect.-
dc.description.statementOfResponsibilityrestriction-
dc.languageEnglish-
dc.publisherSpringer-
dc.relation.isPartOfPHYSICAL AND ENGINEERING SCIENCES IN MEDICINE-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.titleAssessment of dosimetric leaf gap correction factor in Mobius3D commissioning affected by couch top-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Radiation Oncology (방사선종양학교실)-
dc.contributor.googleauthorMin Cheol Han-
dc.contributor.googleauthorJihun Kim-
dc.contributor.googleauthorChae-Seon Hong-
dc.contributor.googleauthorKyung Hwan Chang-
dc.contributor.googleauthorSu Chul Han-
dc.contributor.googleauthorKwangwoo Park-
dc.contributor.googleauthorDong Wook Kim-
dc.contributor.googleauthorMyoung Kyu Park-
dc.contributor.googleauthorYu Yun Noh-
dc.contributor.googleauthorJin Sung Kim-
dc.identifier.doi10.1007/s13246-020-00905-z-
dc.contributor.localIdA05710-
dc.contributor.localIdA05823-
dc.contributor.localIdA04548-
dc.contributor.localIdA01432-
dc.contributor.localIdA05840-
dc.contributor.localIdA05870-
dc.contributor.localIdA05942-
dc.contributor.localIdA05846-
dc.relation.journalcodeJ03946-
dc.identifier.eissn2662-4737-
dc.identifier.pmid32700205-
dc.identifier.urlhttps://link.springer.com/article/10.1007/s13246-020-00905-z-
dc.subject.keywordCouch effect-
dc.subject.keywordDosimetric leaf gap-
dc.subject.keywordMobius-
dc.subject.keywordMobius verification phantom-
dc.contributor.alternativeNameKim, Dong Wook-
dc.contributor.affiliatedAuthor김동욱-
dc.contributor.affiliatedAuthor김지훈-
dc.contributor.affiliatedAuthor김진성-
dc.contributor.affiliatedAuthor박광우-
dc.contributor.affiliatedAuthor장경환-
dc.contributor.affiliatedAuthor한민철-
dc.contributor.affiliatedAuthor한수철-
dc.contributor.affiliatedAuthor홍채선-
dc.citation.volume43-
dc.citation.number3-
dc.citation.startPage1069-
dc.citation.endPage1075-
dc.identifier.bibliographicCitationPHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, Vol.43(3) : 1069-1075, 2020-09-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiation Oncology (방사선종양학교실) > 1. Journal Papers