Anisotropic Total Variation Denoising Technique for Low-Dose Cone-Beam Computed Tomography Imaging

DC Field Value Language
dc.contributor.author이호-
dc.date.accessioned2019-02-12T16:46:37Z-
dc.date.available2019-02-12T16:46:37Z-
dc.date.issued2018-
dc.identifier.issn2508-4445-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/167138-
dc.description.abstractThis study aims to develop an improved Feldkamp-Davis-Kress (FDK) reconstruction algorithm using anisotropic total variation (ATV) minimization to enhance the image quality of low-dose conebeam computed tomography (CBCT). The algorithm first applies a filter that integrates the Shepp- Logan filter into a cosine window function on all projections for impulse noise removal. A total variation objective function with anisotropic penalty is then minimized to enhance the difference between the real structure and noise using the steepest gradient descent optimization with adaptive step sizes. The preserving parameter to adjust the separation between the noise-free and noisy areas is determined by calculating the cumulative distribution function of the gradient magnitude of the filtered image obtained by the application of the filtering operation on each projection. With these minimized ATV projections, voxel-driven backprojection is finally performed to generate the reconstructed images. The performance of the proposed algorithm was evaluated with the catphan503 phantom dataset acquired with the use of a low-dose protocol. Qualitative and quantitative analyses showed that the proposed ATV minimization provides enhanced CBCT reconstruction images compared with those generated by the conventional FDK algorithm, with a higher contrast-to-noise ratio (CNR), lower root-mean-square-error, and higher correlation. The proposed algorithm not only leads to a potential imaging dose reduction in repeated CBCT scans via lower mA levels, but also elicits high CNR values by removing noisy corrupted areas and by avoiding the heavy penalization of striking features.-
dc.description.statementOfResponsibilityopen-
dc.languageEnglish-
dc.publisherKorean Society of Medical Physics-
dc.relation.isPartOfProgress in Medical Physics-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.rightshttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.titleAnisotropic Total Variation Denoising Technique for Low-Dose Cone-Beam Computed Tomography Imaging-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Radiation Oncology (방사선종양학교실)-
dc.contributor.googleauthorHo Lee-
dc.contributor.googleauthorJeongmin Yoon-
dc.contributor.googleauthorEungman Lee-
dc.identifier.doi10.14316/pmp.2018.29.4.150-
dc.contributor.localIdA03323-
dc.relation.journalcodeJ03591-
dc.identifier.eissn2508-4453-
dc.subject.keywordLow-dose CBCT-
dc.subject.keywordFDK-
dc.subject.keywordAnisotropic total variation-
dc.subject.keywordLow mAs-
dc.contributor.alternativeNameLee, Ho-
dc.contributor.affiliatedAuthor이호-
dc.citation.volume29-
dc.citation.number4-
dc.citation.startPage150-
dc.citation.endPage156-
dc.identifier.bibliographicCitationProgress in Medical Physics, Vol.29(4) : 150-156, 2018-
dc.identifier.rimsid58118-
dc.type.rimsART-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiation Oncology (방사선종양학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.