453 707

Cited 0 times in

Practical Implementation of Patient-Specific Quality Assurance for Small and Multiple Brain Tumors in CyberKnife with Fixed Collimators

DC Field Value Language
dc.contributor.author김용배-
dc.contributor.author김진성-
dc.contributor.author박광우-
dc.contributor.author이응만-
dc.contributor.author이호-
dc.date.accessioned2018-08-28T17:28:15Z-
dc.date.available2018-08-28T17:28:15Z-
dc.date.issued2018-
dc.identifier.issn1226-5829-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/162651-
dc.description.abstractThis paper evaluates patient-specific quality assurance (PSQA) in the treatment of small and multiple tumors by the CyberKnife system with fixed collimators, using an ion chamber and EBT3 films. We selected 49 patients with single or multiple brain tumors, and the treatment plans include one to four targets with total volumes ranging from 0.12 cc to 3.74 cc. All PSQA deliveries were performed with a stereotactic dose verification phantom. The A16 microchamber (Standard Imaging, WI, USA) and Gafchromic EBT3 film (Ashland ISP Advanced Materials, NJ, USA) were inserted into the phantom to measure the point dose of the target and the dose distribution, respectively. The film was scanned 1 hr after irradiation by a film digitizer scanner and analyzed using RIT software (Radiological Imaging Technology, CO, USA). The acceptance criteria was <5% for the point dose measurement and >90% gamma passing rate using 3%/3 mm and relative dose difference, respectively. The point dose errors between the calculated and measured dose by the ion chamber were in the range of −17.5% to 8.03%. The mean point dose differences for 5 mm, 7.5 mm, and 10 mm fixed cone size was −11.1%, −4.1%, and −1.5%, respectively. The mean gamma passing rates for all cases was 96.1%. Although the maximum dose distribution of multiple targets was not shown in the film, gamma distribution showed that dose verification for multiple tumors can be performed. The use of the microchamber and EBT3 film made it possible to verify the dosimetric and mechanical accuracy of small and multiple targets. In particular, the correction factors should be applied to small fixed collimators less than 10 mm.-
dc.description.statementOfResponsibilityopen-
dc.languageKorean, English-
dc.publisher한국의학물리학회-
dc.relation.isPartOfKorean Journal of Medical Physics (의학물리)-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.rightshttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.titlePractical Implementation of Patient-Specific Quality Assurance for Small and Multiple Brain Tumors in CyberKnife with Fixed Collimators-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine-
dc.contributor.departmentDept. of Radiation Oncology-
dc.contributor.googleauthorEungman Lee-
dc.contributor.googleauthorKwangwoo Park-
dc.contributor.googleauthorJin Sung Kim-
dc.contributor.googleauthorYong Bae Kim-
dc.contributor.googleauthorHo Lee-
dc.identifier.doi10.14316/pmp.2018.29.2.53-
dc.contributor.localIdA00744-
dc.contributor.localIdA04548-
dc.contributor.localIdA01432-
dc.contributor.localIdA05009-
dc.contributor.localIdA03323-
dc.relation.journalcodeJ02057-
dc.identifier.eissn2288-9620-
dc.subject.keywordCyberKnife-
dc.subject.keywordMicrochamber-
dc.subject.keywordEBT3-
dc.subject.keywordPatient-specific quality assurance-
dc.subject.keywordBrain tumor-
dc.contributor.alternativeNameKim, Yong Bae-
dc.contributor.alternativeNameKim, Jinsung-
dc.contributor.alternativeNamePark, Kwang Woo-
dc.contributor.alternativeNameLee, Eung Man-
dc.contributor.alternativeNameLee, Ho-
dc.contributor.affiliatedAuthorKim, Yong Bae-
dc.contributor.affiliatedAuthorKim, Jinsung-
dc.contributor.affiliatedAuthorPark, Kwang Woo-
dc.contributor.affiliatedAuthorLee, Eung Man-
dc.contributor.affiliatedAuthorLee, Ho-
dc.citation.volume29-
dc.citation.number2-
dc.citation.startPage53-
dc.citation.endPage58-
dc.identifier.bibliographicCitationKorean Journal of Medical Physics (의학물리), Vol.29(2) : 53-58, 2018-