dc.description.abstractPURPOSE: The self-pressurized air-Q(R) (air-Q SP) intubating laryngeal airway is a relatively new supraglottic airway (SGA) device. The intracuff pressure of air-Q dynamically equilibrates with the airway pressure and adjusts to the patient's pharyngeal and periglottic anatomy, potentially providing improved airway fit and seal. The aim of this prospective randomized study was to compare the clinical performance of air-Q to the LMA(R) Classic SGA. METHODS: Adult patients requiring general anesthesia for elective surgery were prospectively enrolled and randomly assigned to either air-Q SP or the LMA Classic SGA. Oropharyngeal leak pressure (primary endpoint), success rate, insertion features (insertion time, ease of insertion, requirement for device manipulation), sealing function, gastric insufflation, bronchoscopic view, and oropharyngeal complications at device insertion and following its removal (sore throat, dysphagia, dysphonia) were compared. RESULTS: The mean (standard deviation [SD]) oropharyngeal leak pressure just after insertion was similar in the air-Q SP and LMA [16.8 (4.9) vs 18.6 (5.5) cm H2O, respectively; mean difference, 1.8 cm H2O; 95% CI, -0.5 to 4.2; P = 0.13] and did not differ at ten minutes following device insertion. Median [interquartile range (IQR)] peak inspiratory pressure just after insertion was lower in the air-Q SP (11.0 [10.0-13.0] vs 13.0 [11.0-14.0] cmH2O, median difference, 1.0 cm H2O; 95% CI, 0.0 to 2.0; P = 0.03) but no difference was observed at ten minutes. The median [IQR] insertion time was faster with the air-Q SP (15.9 [13.6-20.3] sec vs 24 [21.2-27.1] sec; median difference, 8.1 sec; 95% CI, 5.6 to 9.9; P < 0.001) and improved bronchoscopic viewing grade were seen with the air-Q SP immediately after insertion (P < 0.001). No differences between the groups were observed with respect to the rate of successful insertion at first attempt, overall insertion success rate, ease of insertion, and complications. CONCLUSIONS: The air-Q SP had similar leak pressures but a faster insertion time and superior bronchoscopic viewing grade when compared with the LMA Classic. The air-Q SP is a suitable alternative to the LMA Classic in adult patients and may be a superior conduit for tracheal intubation. TRIAL REGISTRATION: www.clinicaltrials.gov (NCT02206438). Registered 1 August 2014.-
dc.description.statementOfResponsibilityrestriction-
dc.languageEnglish-
dc.publisherSpringer New York-
dc.relation.isPartOfCANADIAN JOURNAL OF ANESTHESIA-JOURNAL CANADIEN D ANESTHESIE-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.rightshttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.titleSelf-pressurized air-Q((R)) intubating laryngeal airway versus the LMA((R)) Classic: a randomized clinical trial-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine-
dc.contributor.departmentDept. of Anesthesiology and Pain Medicine-
dc.contributor.googleauthorSang Hee Ha-
dc.contributor.googleauthorMin-Soo Kim-
dc.contributor.googleauthorJiwoo Suh-
dc.contributor.googleauthorJong Seok Lee-
dc.identifier.doi10.1007/s12630-018-1082-6-
dc.contributor.localIdA00463-
dc.contributor.localIdA04584-
dc.contributor.localIdA03141-
dc.relation.journalcodeJ00426-
dc.identifier.eissn1496-8975-
dc.identifier.pmid29524199-
dc.identifier.urlhttps://link.springer.com/article/10.1007%2Fs12630-018-1082-6-
dc.contributor.alternativeNameKim, Min Soo-
dc.contributor.alternativeNameSeo, Jiwoo-
dc.contributor.alternativeNameLee, Jong Seok-
dc.contributor.affiliatedAuthorKim, Min Soo-
dc.contributor.affiliatedAuthorSeo, Jiwoo-
dc.contributor.affiliatedAuthorLee, Jong Seok-
dc.citation.volume65-
dc.citation.number5-
dc.citation.startPage543-
dc.citation.endPage550-
dc.identifier.bibliographicCitationCANADIAN JOURNAL OF ANESTHESIA-JOURNAL CANADIEN D ANESTHESIE, Vol.65(5) : 543-550, 2018-
dc.identifier.rimsid59851-
dc.type.rimsART-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Anesthesiology and Pain Medicine (마취통증의학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.