253 516

Cited 0 times in

The Results of Long-term Follow-up of Total Hip Arthroplasty Using Hydroxyapatite-coated Cups

DC Field Value Language
dc.contributor.author박관규-
dc.contributor.author신근영-
dc.contributor.author양익환-
dc.contributor.author이우석-
dc.contributor.author이현희-
dc.contributor.author한창동-
dc.date.accessioned2018-03-26T17:07:15Z-
dc.date.available2018-03-26T17:07:15Z-
dc.date.issued2015-
dc.identifier.issn2287-3260-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/157214-
dc.description.abstractPURPOSE: The aim of this study was to report the long-term outcome and the failure mechanism of cementless total hip arthroplasty (THA) using hydroxyapatite (HA)-coated acetabular cup. MATERIALS AND METHODS: From January 1992 to May 1994, a total of 123 consecutive cementless primary THAs were performed using a HA-coated acetabular cup with metal-on-polyethylene articulation. We retrospectively evaluated 66 hips available for follow-up at a mean 18.3 years (range, 10.4-23.6 years). The survival analysis was performed by the Kaplan-Meier method. We defined end point as any failure that required a reoperation of acetabular component. RESULTS: Thirty-nine of 66 hips (59.1%) were defined as a failure for progressive acetabular osteolysis or aseptic loosening of the cup. Acetabular osteolysis was observed in 47 hips (71.2%) and 33 hips (50.0%) were revised because of cup loosening. The Kaplan-Meier method showed the survival rate of the acetabular cup to be 46.3% at 15 years and 34.8% at 20 years for any failure that required a reoperation of acetabular component. CONCLUSION: The long-term survival rate of THA using HA-coated acetabular cup was unsatisfactory, and it was attributed to vulnerable property of HA coating and progressive osteolysis.-
dc.description.statementOfResponsibilityopen-
dc.languageEnglish-
dc.publisher대한고관절학회-
dc.relation.isPartOfHip & Pelvis-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.rightshttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.titleThe Results of Long-term Follow-up of Total Hip Arthroplasty Using Hydroxyapatite-coated Cups-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine-
dc.contributor.departmentDept. of Orthopedic Surgery-
dc.contributor.googleauthorChang-Dong Han-
dc.contributor.googleauthorKeun-Young Shin-
dc.contributor.googleauthorHyun-Hee Lee-
dc.contributor.googleauthorKwan-Kyu Park-
dc.contributor.googleauthorIck-Hwan Yang-
dc.contributor.googleauthorWoo-Suk Lee-
dc.identifier.doi10.5371/hp.2015.27.4.209-
dc.contributor.localIdA01428-
dc.contributor.localIdA04764-
dc.contributor.localIdA02313-
dc.contributor.localIdA02992-
dc.contributor.localIdA04646-
dc.contributor.localIdA04330-
dc.relation.journalcodeJ00989-
dc.identifier.eissn2287-3279-
dc.identifier.pmid27536628-
dc.subject.keywordAcetabular cup-
dc.subject.keywordHydroxyapatite-coated-
dc.subject.keywordLong-term outcome-
dc.subject.keywordTotal hip arthroplasty-
dc.contributor.alternativeNamePark, Kwan Kyu-
dc.contributor.alternativeNameShin, Keun-Young-
dc.contributor.alternativeNameYang, Ick Hwan-
dc.contributor.alternativeNameLee, Woo Suk-
dc.contributor.alternativeNameLee, Hyun Hee-
dc.contributor.alternativeNameHan, Chang Dong-
dc.contributor.affiliatedAuthorPark, Kwan Kyu-
dc.contributor.affiliatedAuthorShin, Keun-Young-
dc.contributor.affiliatedAuthorYang, Ick Hwan-
dc.contributor.affiliatedAuthorLee, Woo Suk-
dc.contributor.affiliatedAuthorLee, Hyun Hee-
dc.contributor.affiliatedAuthorHan, Chang Dong-
dc.citation.volume27-
dc.citation.number4-
dc.citation.startPage209-
dc.citation.endPage215-
dc.identifier.bibliographicCitationHip & Pelvis, Vol.27(4) : 209-215, 2015-
dc.identifier.rimsid41774-
dc.type.rimsART-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Orthopedic Surgery (정형외과학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.