0 479

Cited 54 times in

Low-interference washing-free electrochemical immunosensor using glycerol-3-phosphate dehydrogenase as an enzyme label

Authors
 Gorachand Dutta,† Seonhwa Park,† Amardeep Singh,† Jeongwook Seo,† Sinyoung Kim  ;  Haesik Yang 
Citation
 ANALYTICAL CHEMISTRY, Vol.87(7) : 3574-3578, 2015 
Journal Title
ANALYTICAL CHEMISTRY
ISSN
 0003-2700 
Issue Date
2015
MeSH
Antibodies, Immobilized/chemistry ; Biosensing Techniques/methods ; Electrochemical Techniques/methods* ; Electrodes ; Glycerolphosphate Dehydrogenase ; Humans ; Immunoenzyme Techniques/methods* ; Limit of Detection ; Tin Compounds/chemistry ; Troponin I/analysis ; Troponin I/blood*
Abstract
In washing-free electrochemical detection, various redox and reactive species cause significant interference. To minimize this interference, we report a washing-free electrochemical immunosensor using flavin adenine dinucleotide (FAD)-dependent glycerol-3-phosphate dehydrogenase (GPDH) and glycerol-3-phosphate (GP) as an enzyme label and its substrate, respectively, because the reaction of FAD-dependent dehydrogenases with dissolved O2 is slow and the level of GP preexisting in blood is low (<0.1 mM). A combination of a low electrocatalytic indium-tin oxide (ITO) electrode and fast electron-mediating Ru(NH3)6(3+) is employed to obtain a high signal-to-background ratio via proximity-dependent electron mediation of Ru(NH3)6(3+) between the ITO electrode and the GPDH label. Electrochemical oxidation of GPDH-generated Ru(NH3)6(2+) is performed at 0.05 V vs Ag/AgCl, at which point the electrochemical interference is very low. When a washing-free immunosensor is applied to cardiac troponin I detection in human serum, the calculated detection limit is approximately 10 pg/mL, indicating that the immunosensor is very sensitive in spite of the use of washing-free detection with a short detection period (10 min for incubation and 100 s for electrochemical measurement). The low-interference washing-free electrochemical immunosensor shows good promise for fast and simple point-of-care testing.
Full Text
http://pubs.acs.org/doi/10.1021/ac504485a
DOI
10.1021/ac504485a
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Laboratory Medicine (진단검사의학교실) > 1. Journal Papers
Yonsei Authors
Kim, Sin Young(김신영) ORCID logo https://orcid.org/0000-0002-2609-8945
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/156743
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links