1130 884

Cited 0 times in

자동 요분석기 CLINITEK Novus의 성능 평가

DC Field Value Language
dc.contributor.author김정호-
dc.contributor.author원영철-
dc.contributor.author이상국-
dc.contributor.author임지숙-
dc.contributor.author조선미-
dc.date.accessioned2017-10-26T07:16:29Z-
dc.date.available2017-10-26T07:16:29Z-
dc.date.issued2016-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/151823-
dc.description.abstractBackground: We aimed to evaluate the performance of the CLINITEK Novus urine chemistry analyzer (Siemens, UK). Methods: The precision, correlation, and carryover study were performed using two kinds of commercial quality control materials and 40-55 freshly collected patient specimens. We calculated exact and within-1-block agreement, along with kappa agreement, to compare the semi-quantitative results between urine chemistry analyzers. The urine specific gravity taken by a refractometer was compared with the analyzer results. Moreover, we analyzed additional urine specimens for protein to evaluate the agreement of results between those of the CLINITEK Novus and the AU680 analyzers (Beckman Coulter, Japan). Results: The precision study showed acceptable results; within-1-block agreement was 100% in all tested items. The urine chemistry results from the CLNITEK Novus analyzer demonstrated ≥85.1% within-1-block agreements with those of the Uriscan Super, and the kappa test results were ≥0.81. The comparison of specific gravity with manual refractometer showed a good correlation (r=0.991), and the protein comparison with the AU680 analyzer also showed a good correlation (with exact and within-1-block agreements being 75.9% and 100.0%, respectively). The carryover rates were 0% in all tested items, except specific gravity and heavy blood tests. Conclusions: The CLINITEK Novus analyzer showed good performance in terms of precision, comparison, and carryover in this study. Therefore, the CLINITEK Novus automated urine analysis is expected to be useful for routine urinalysis in a clinical laboratory.-
dc.description.statementOfResponsibilityopen-
dc.languageKorean-
dc.publisherKorean Society for Laboratory Medicine-
dc.relation.isPartOfLaboratory Medicine Online-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.title자동 요분석기 CLINITEK Novus의 성능 평가-
dc.title.alternativePerformance Evaluation of the CLINITEK Novus Automated Urine Chemistry Analyzer-
dc.typeArticle-
dc.publisher.locationKorea (South)-
dc.contributor.collegeCollege of Medicine-
dc.contributor.departmentDept. of Laboratory Medicine-
dc.contributor.googleauthor임지숙-
dc.contributor.googleauthor이상국-
dc.contributor.googleauthor조선미-
dc.contributor.googleauthor원영철-
dc.contributor.googleauthor김정호-
dc.identifier.doi10.3343/lmo.2016.6.3.147-
dc.contributor.localIdA04983-
dc.contributor.localIdA02810-
dc.contributor.localIdA04655-
dc.contributor.localIdA03827-
dc.contributor.localIdA00903-
dc.relation.journalcodeJ02151-
dc.identifier.eissn2093-6338-
dc.subject.keywordPerformance evaluation-
dc.subject.keywordAutomated urinalysis-
dc.subject.keywordCLINITEK Novus-
dc.contributor.alternativeNameKim, Jeong Ho-
dc.contributor.alternativeNameWon, Young Cheol-
dc.contributor.alternativeNameLee, Sang Guk-
dc.contributor.alternativeNameYim, Jisook-
dc.contributor.alternativeNameCho, Sun Mi-
dc.contributor.affiliatedAuthorWon, Young Cheol-
dc.contributor.affiliatedAuthorLee, Sang Guk-
dc.contributor.affiliatedAuthorYim, Jisook-
dc.contributor.affiliatedAuthorCho, Sun Mi-
dc.contributor.affiliatedAuthorKim, Jeong Ho-
dc.citation.volume6-
dc.citation.number3-
dc.citation.startPage147-
dc.citation.endPage151-
dc.identifier.bibliographicCitationLaboratory Medicine Online, Vol.6(3) : 147-151, 2016-
dc.date.modified2017-10-24-
dc.identifier.rimsid45832-
dc.type.rimsART-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Laboratory Medicine (진단검사의학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.