0 917

Cited 5 times in

A blocking-wire technique for removal of a broken hollow intramedullary nail.

DC Field Value Language
dc.contributor.author강동현-
dc.contributor.author박영창-
dc.contributor.author석상옥-
dc.contributor.author양규현-
dc.contributor.author윤여권-
dc.date.accessioned2017-02-27T08:18:14Z-
dc.date.available2017-02-27T08:18:14Z-
dc.date.issued2016-
dc.identifier.issn0020-1383-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/147147-
dc.description.abstractWe present a simple method to remove the distal portion of the broken nail just using the bulb-tipped guide pin and a blocking Kirschner wire. At first, we removed the proximal part of the broken nail and all interlocking screws. Next, we introduced the bulb-tipped guide wire into the distal part of the nail with fluoroscopic guidance until it passed the interlocking hole that would be used. After snugly fitting the blocking Kirschner wire into the interlocking hole of the nail, the bulb-tipped guide wire is then pulled to engage the blocking wire complex in the interlocking hole. The bulb-tipped guide wire and broken nail are removed by sequential stroke using a ram.-
dc.description.statementOfResponsibilityrestriction-
dc.format.extent1601~1603-
dc.languageEnglish-
dc.publisherElsevier-
dc.relation.isPartOfINJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.subject.MESHBone Nails*-
dc.subject.MESHBone Wires*-
dc.subject.MESHDevice Removal/instrumentation*-
dc.subject.MESHDevice Removal/methods*-
dc.subject.MESHEquipment Failure-
dc.subject.MESHFluoroscopy-
dc.subject.MESHFracture Fixation, Intramedullary/adverse effects*-
dc.subject.MESHHumans-
dc.titleA blocking-wire technique for removal of a broken hollow intramedullary nail.-
dc.typeArticle-
dc.publisher.locationNetherlands-
dc.contributor.collegeCollege of Medicine-
dc.contributor.departmentDept. of Orthopedic Surgery-
dc.contributor.googleauthorXuan-Lin Zheng-
dc.contributor.googleauthorYoung-Chang Park-
dc.contributor.googleauthorDong-Hyun Kang-
dc.contributor.googleauthorSang-Ok Seok-
dc.contributor.googleauthorYeo-Kwon Yoon-
dc.contributor.googleauthorKyu-Hyun Yang-
dc.identifier.doi10.1016/j.injury.2016.04.039-
dc.contributor.localIdA00014-
dc.contributor.localIdA04569-
dc.contributor.localIdA04586-
dc.contributor.localIdA02282-
dc.contributor.localIdA04614-
dc.relation.journalcodeJ01064-
dc.identifier.eissn1879-0267-
dc.identifier.pmid27170540-
dc.identifier.urlhttp://www.sciencedirect.com/science/article/pii/S0020138316301462-
dc.subject.keywordBroken nail-
dc.subject.keywordNarrow hollow-
dc.subject.keywordRemoval-
dc.contributor.alternativeNameKang, Dong Hyun-
dc.contributor.alternativeNamePark, Young Chang-
dc.contributor.alternativeNameSeok, Sang Ok-
dc.contributor.alternativeNameYang, Kyu Hyun-
dc.contributor.alternativeNameYoon, Yeo Kwon-
dc.contributor.affiliatedAuthorKang, Dong Hyun-
dc.contributor.affiliatedAuthorPark, Young Chang-
dc.contributor.affiliatedAuthorSeok, Sang Ok-
dc.contributor.affiliatedAuthorYang, Kyu Hyun-
dc.contributor.affiliatedAuthorYoon, Yeo Kwon-
dc.citation.volume47-
dc.citation.number7-
dc.citation.startPage1601-
dc.citation.endPage1603-
dc.identifier.bibliographicCitationINJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, Vol.47(7) : 1601-1603, 2016-
dc.date.modified2017-02-24-
dc.identifier.rimsid47178-
dc.type.rimsART-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Orthopedic Surgery (정형외과학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.