Cited 0 times in
폐전절제술을 받아 기관이 오른쪽으로 편향되어 있는 환자에서 경동맥 내막절제술 중에 발생한 강화 기관내관의 부분폐쇄 : 증례 보고
DC Field | Value | Language |
---|---|---|
dc.contributor.author | 배선준 | - |
dc.date.accessioned | 2016-02-19T11:16:01Z | - |
dc.date.available | 2016-02-19T11:16:01Z | - |
dc.date.issued | 2001 | - |
dc.identifier.issn | 0302-5780 | - |
dc.identifier.uri | https://ir.ymlib.yonsei.ac.kr/handle/22282913/142764 | - |
dc.description.abstract | Endotracheal tube obstruction during anesthesia can have many causes. Hyperinflation of the remaining lung after a pneumonectomy can severely displace the trachea, and attachment of an endotracheal tube tip to the wall of a deviated trachea may also cause severe airway obstruction. The right lung of the patient was removed 3 years ago due to lung cancer. Compensatory hyperinflation of the left lung and severe right-sided tracheal deviation was seen on a chest X-ray. An armored endotracheal tube without Murphy's eye was used. Two hours after beginning the operation, peak airway pressure and PETCO2 began to increase gradually. A wheezing-like sound was heard. Bronchospasm was suspected, but signs of a spasm were not relieved by medications. The signs completely disappeared after pulling the tube 2 cm proximal. The position of the tube should be confirmed by fiberoptic bronchoscopy or chest X-ray after intubation when the trachea is deviated. | - |
dc.description.statementOfResponsibility | open | - |
dc.format.extent | 105~109 | - |
dc.relation.isPartOf | Journal of Korean Society of Anesthesiologist (대한마취과학회지) | - |
dc.rights | CC BY-NC-ND 2.0 KR | - |
dc.rights.uri | https://creativecommons.org/licenses/by-nc-nd/2.0/kr/ | - |
dc.title | 폐전절제술을 받아 기관이 오른쪽으로 편향되어 있는 환자에서 경동맥 내막절제술 중에 발생한 강화 기관내관의 부분폐쇄 : 증례 보고 | - |
dc.title.alternative | Partial Obstruction of an Armored Endotracheal Tube during a Carotid Endarterectomy due to Tracheal Deviation in a Pnemonectomized Patient | - |
dc.type | Article | - |
dc.contributor.college | College of Medicine (의과대학) | - |
dc.contributor.department | Dept. of Anesthesiology (마취통증의학) | - |
dc.contributor.googleauthor | 배선준 | - |
dc.contributor.googleauthor | 김기준 | - |
dc.contributor.googleauthor | 김종훈 | - |
dc.contributor.googleauthor | 김건호 | - |
dc.contributor.googleauthor | 박윤곤 | - |
dc.admin.author | false | - |
dc.admin.mapping | false | - |
dc.contributor.localId | A01795 | - |
dc.relation.journalcode | J01540 | - |
dc.subject.keyword | Complications | - |
dc.subject.keyword | airway | - |
dc.subject.keyword | endotracheal tube obstruction | - |
dc.subject.keyword | Equipment | - |
dc.subject.keyword | tubes | - |
dc.subject.keyword | endotracheal | - |
dc.contributor.alternativeName | Bai, Sun Joon | - |
dc.contributor.affiliatedAuthor | Bai, Sun Joon | - |
dc.rights.accessRights | free | - |
dc.citation.volume | 41 | - |
dc.citation.number | 1 | - |
dc.citation.startPage | 105 | - |
dc.citation.endPage | 109 | - |
dc.identifier.bibliographicCitation | Journal of Korean Society of Anesthesiologist (대한마취과학회지), Vol.41(1) : 105-109, 2001 | - |
dc.identifier.rimsid | 29830 | - |
dc.type.rims | ART | - |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.