263 658

Cited 31 times in

Evaluation of VITEK mass spectrometry (MS), a matrix-assisted laser desorption ionization time-of-flight MS system for identification of anaerobic bacteria

Authors
 Wonmok Lee  ;  Myungsook Kim  ;  Dongeun Yong  ;  Seok Hoon Jeong  ;  Kyungwon Lee  ;  Yunsop Chong 
Citation
 ANNALS OF LABORATORY MEDICINE, Vol.35(1) : 69-75, 2015 
Journal Title
ANNALS OF LABORATORY MEDICINE
ISSN
 2234-3806 
Issue Date
2015
MeSH
Bacteria, Anaerobic/genetics* ; Bacteria, Anaerobic/isolation & purification ; Bacterial Typing Techniques/instrumentation* ; Bacterial Typing Techniques/methods* ; Body Fluids/microbiology ; Databases, Genetic ; Humans ; RNA, Ribosomal, 16S/analysis* ; RNA, Ribosomal, 16S/metabolism ; Sequence Analysis, DNA ; Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization*
Keywords
Anaerobic bacteria ; MALDI-TOF MS ; VITEK MS
Abstract
BACKGROUND: By conventional methods, the identification of anaerobic bacteria is more time consuming and requires more expertise than the identification of aerobic bacteria. Although the matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) systems are relatively less studied, they have been reported to be a promising method for the identification of anaerobes. We evaluated the performance of the VITEK MS in vitro diagnostic (IVD; 1.1 database; bioMérieux, France) in the identification of anaerobes.

METHODS: We used 274 anaerobic bacteria isolated from various clinical specimens. The results for the identification of the bacteria by VITEK MS were compared to those obtained by phenotypic methods and 16S rRNA gene sequencing.

RESULTS: Among the 249 isolates included in the IVD database, the VITEK MS correctly identified 209 (83.9%) isolates to the species level and an additional 18 (7.2%) at the genus level. In particular, the VITEK MS correctly identified clinically relevant and frequently isolated anaerobic bacteria to the species level. The remaining 22 isolates (8.8%) were either not identified or misidentified. The VITEK MS could not identify the 25 isolates absent from the IVD database to the species level.

CONCLUSIONS: The VITEK MS showed reliable identifications for clinically relevant anaerobic bacteria.
Files in This Item:
T201500081.pdf Download
DOI
10.3343/alm.2015.35.1.69
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Laboratory Medicine (진단검사의학교실) > 1. Journal Papers
Yonsei Authors
Yong, Dong Eun(용동은) ORCID logo https://orcid.org/0000-0002-1225-8477
Lee, Kyungwon(이경원) ORCID logo https://orcid.org/0000-0003-3788-2134
Jeong, Seok Hoon(정석훈) ORCID logo https://orcid.org/0000-0001-9290-897X
Chong, Yun Sop(정윤섭)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/139212
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links