528 448

Cited 0 times in

국산 초저온 냉동보관 혈액백의 평가

Other Titles
 Evaluation of Domestic Cryostorage Blood Bags. 
Authors
 김현옥  ;  김주원  ;  김홍련 
Citation
 Korean Journal of Blood Transfusion (대한수혈학회지), Vol.17(1) : 48-53, 2006 
Journal Title
Korean Journal of Blood Transfusion(대한수혈학회지)
ISSN
 1226-9336 
Issue Date
2006
Keywords
Cryostorage blood bag ; Cryosolution ; Freezing condition
Abstract
Background: Recent studies have shown that the optimal temperature at which the cord blood and the peripheral blood stem cells need to be cryopreserved is below -160oC in a nitrogen tank. This study evaluated a domestic cryostorage cord blood processing bag (CBP set) made from poly (ethylene-co-vinylacetate) (EVA) recently produced by the Medi-Rution Company in Korea.
Methods: The domestic cryostorage bags was evaluated using 100 bags from Medi-Rution, which were divided into 3 different groups and compared with 20 bags produced by the Thermogenesis company from the United States as a control. In the first group, the mononuclear cells separated from the human blood sample were divided equally into the cryosolution using DMSO, auto plasma and RPMI media. The blood was preserved in 20 domestic bags and 20 bags produced by the Thermogenesis company. In the second group, the cells were stored in the cryosolution with either auto plasma or fetal bovine serum. In the third group, 20 bags were frozen at -70oC before being stored in the nitrogen tank whereas the other 20 bags were frozen in a controlled refrigerator before being stored in the nitrogen tank at -196oC. The bags containing the peripheral blood cells were stored in the nitrogen tank at vapor phase for a period of one year and evaluated for the rate of breakage during thawing, recovery rates and viability of the cell after thawing.
Results: None of the bags evaluated in this study showed any signs of breakage or leakage of the products. All three groups tested showed similar results regardless of where the bags were made, the composition of the cryosolution or the early freezing conditions.
Conclusion: The CBP set cryostorage bags recently produced by Medi-Rution can be used to preserve blood cells at extremely low temperatures for a one year period without any significant breakage.
Files in This Item:
T200600426.pdf Download
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Laboratory Medicine (진단검사의학교실) > 1. Journal Papers
Yonsei Authors
Kim, Hyun Ok(김현옥) ORCID logo https://orcid.org/0000-0002-4964-1963
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/109278