476 434

Cited 0 times in

두종의external hex implant의 변연골 흡수에 관한 연구 : 예비연구 (preliminary study)

DC Field Value Language
dc.contributor.author한종현-
dc.date.accessioned2015-05-19T16:46:17Z-
dc.date.available2015-05-19T16:46:17Z-
dc.date.issued2008-
dc.identifier.issn0301-2875-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/106914-
dc.description.abstractStatement of problem: Changes of the marginal bone around dental implants have significance not only for the functional maintenance but also for the esthetic success of the implant. It was proposed that bone-retention elements such as microthreads at the coronal part of implant might help maintain the marginal bone level. Purpose: This study was designed to evaluate the effect of microthread configuration within the marginal coronal portion of the implant fixture at the marginal bone changes after loading around two different external hex implants. Material and methods: Twenty-four patients were included and randomly assigned to treatment with Branemark system implants (Group 1, rough-surfaced implants, n=20) and Oneplant system implants (Group 2, rough-surfaced neck with microthreads, n=20). Clinical and radiographic examinations were conducted at baseline (implant loading) and 1 year postloading. Data analysis was performed by the SAS statistical package version 9.1.3 (SAS Institute, Cary, NC, USA) and the final model was calculated by the MIXED procedure (three-level ANCOVA) for marginal bone change of each test group at baseline and 1 year follow-up. Results: Comparing to baseline, significant differences were noted in marginal bone level changes for the 2 groups at 1 year follow-up (P<0.05). Group 1 had a mean crestal bone level changes of $0.83{\pm}0.31mm$; Group 2 had a mean crestal bone level changes of $0.44{\pm}0.36mm$. Rough-surfaced with microthreads implants showed significantly less marginal bone loss than rough surfaced neck without microthread implants. Conclusion: A rough surface with microthreads at the implant was beneficial design to maintain the marginal bone level against functional loading.-
dc.description.statementOfResponsibilityopen-
dc.format.extent169~174-
dc.relation.isPartOfJournal of Korean Academy of Prosthodontics (대한치과보철학회지)-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.title두종의external hex implant의 변연골 흡수에 관한 연구 : 예비연구 (preliminary study)-
dc.title.alternativeRadiographic evaluation of marginal bone resorption around two types of external hex implants : preliminary study-
dc.typeArticle-
dc.contributor.collegeCollege of Dentistry (치과대학)-
dc.contributor.departmentDept. of Prosthodontics (보철과학)-
dc.contributor.googleauthor이지은-
dc.contributor.googleauthor허성주-
dc.contributor.googleauthor곽재영-
dc.contributor.googleauthor김성균-
dc.contributor.googleauthor한종현-
dc.admin.authorfalse-
dc.admin.mappingfalse-
dc.contributor.localIdA04326-
dc.relation.journalcodeJ01493-
dc.identifier.pmidMarginal bone level ; Microthreads ; Rough surface-
dc.subject.keywordMarginal bone level-
dc.subject.keywordMicrothreads-
dc.subject.keywordRough surface-
dc.contributor.alternativeNameHan, Jong Hyoun-
dc.contributor.affiliatedAuthorHan, Jong Hyoun-
dc.rights.accessRightsfree-
dc.citation.volume46-
dc.citation.number2-
dc.citation.startPage169-
dc.citation.endPage174-
dc.identifier.bibliographicCitationJournal of Korean Academy of Prosthodontics (대한치과보철학회지), Vol.46(2) : 169-174, 2008-
Appears in Collections:
2. College of Dentistry (치과대학) > Dept. of Prosthodontics (보철과학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.