2 432

Cited 12 times in

Carbon nanotube-based biosensor for detection of matrix metallopeptidase-9 and S-100B

Authors
 Hyoung-Sub Lee  ;  Je-Seung Oh  ;  Young-Wook Chang  ;  Yoon-Jung Park  ;  Jeon-Soo Shin  ;  Kyung-Hwa Yoo 
Citation
 CURRENT APPLIED PHYSICS, Vol.9(4 suppl) : 270-272, 2009 
Journal Title
CURRENT APPLIED PHYSICS
ISSN
 1567-1739 
Issue Date
2009
Keywords
CNT–FET ; Biosensor ; MMP-9 ; S-100B
Abstract
We fabricated carbon nanotube (CNT)-based biosensor with field effect transistor structure for detection of matrix metallopeptidase-9 (MMP-9) and S-100B. To minimize contact effects, the metal electrodes were passivated with SiOx film, and anti-MMP-9 and S-100B antibody were immobilized on the CNT surface using a chemical linker. When MMP-9 or S-100B antigen was added to bind anti-MMP-9 or S-100B antibody, the decrease in electrical conductance was observed. We have also investigated the dependence of MMP-9 and S-100B concentration on the conductance change. A linear relationship was found between the conductance and the concentration in semi-logarithm scale.
Full Text
http://www.sciencedirect.com/science/article/pii/S1567173909003514
DOI
10.1016/j.cap.2009.07.003
Appears in Collections: