Cited 42 times in
Analysis of biomechanical changes after removal of instrumentation in lumbar arthrodesis by finite element analysis.
DC Field | Value | Language |
---|---|---|
dc.contributor.author | 김학선 | - |
dc.contributor.author | 김호중 | - |
dc.contributor.author | 문성환 | - |
dc.contributor.author | 문은수 | - |
dc.contributor.author | 박진오 | - |
dc.contributor.author | 이환모 | - |
dc.date.accessioned | 2015-04-23T17:13:04Z | - |
dc.date.available | 2015-04-23T17:13:04Z | - |
dc.date.issued | 2010 | - |
dc.identifier.issn | 0140-0118 | - |
dc.identifier.uri | https://ir.ymlib.yonsei.ac.kr/handle/22282913/102045 | - |
dc.description.abstract | The purpose of this study is to investigate the change in biomechanical milieu following removal of pedicle screws in instrumented single level lumbar arthrodesis. Using a validated finite element (FE) model of the intact lumbar spine (L2-5), two scenarios of L3-4 lumbar fusion were simulated: posterolateral fusion (PLF) at L3-4 using pedicle screws (PLF with pedicle screws; WiP) and L3-4 lumbar posterolateral fusion state after removal of pedicle screws (PLF without pedicle screws; WoP). The WiP model had greater range of motion (ROM) at each adjacent segment than the WoP model. This phenomenon became pronounced at the proximal adjacent segment under flexion moment. Similarly, removal of pedicle screws (the WoP model) relieved the maximal von Mises stress at adjacent segments under 4 moments compared to the WiP model. This study demonstrated that removal of pedicle screws could decrease stiffness of fusion segments, which would reduce the disk stress of adjacent segments | - |
dc.description.statementOfResponsibility | open | - |
dc.format.extent | 703~709 | - |
dc.relation.isPartOf | MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING | - |
dc.rights | CC BY-NC-ND 2.0 KR | - |
dc.rights.uri | https://creativecommons.org/licenses/by-nc-nd/2.0/kr/ | - |
dc.subject.MESH | Biomechanical Phenomena | - |
dc.subject.MESH | Bone Screws | - |
dc.subject.MESH | Device Removal | - |
dc.subject.MESH | Finite Element Analysis | - |
dc.subject.MESH | Humans | - |
dc.subject.MESH | Lumbar Vertebrae/physiopathology | - |
dc.subject.MESH | Lumbar Vertebrae/surgery* | - |
dc.subject.MESH | Male | - |
dc.subject.MESH | Middle Aged | - |
dc.subject.MESH | Models, Biological* | - |
dc.subject.MESH | Range of Motion, Articular | - |
dc.subject.MESH | Spinal Fusion/instrumentation* | - |
dc.title | Analysis of biomechanical changes after removal of instrumentation in lumbar arthrodesis by finite element analysis. | - |
dc.type | Article | - |
dc.contributor.college | College of Medicine (의과대학) | - |
dc.contributor.department | Dept. of Orthopedic Surgery (정형외과학) | - |
dc.contributor.googleauthor | Ho-Joong Kim | - |
dc.contributor.googleauthor | Heoung-Jae Chun | - |
dc.contributor.googleauthor | Seong-Hwan Moon | - |
dc.contributor.googleauthor | Kyoung-Tak Kang | - |
dc.contributor.googleauthor | Hak-Sun Kim | - |
dc.contributor.googleauthor | Jin-Oh Park | - |
dc.contributor.googleauthor | Eun-Su Moon | - |
dc.contributor.googleauthor | Joon-Seok Sohn | - |
dc.contributor.googleauthor | Hwan-Mo Lee | - |
dc.identifier.doi | 10.1007/s11517-010-0621-2 | - |
dc.admin.author | false | - |
dc.admin.mapping | false | - |
dc.contributor.localId | A01186 | - |
dc.contributor.localId | A01093 | - |
dc.contributor.localId | A01365 | - |
dc.contributor.localId | A01372 | - |
dc.contributor.localId | A01703 | - |
dc.contributor.localId | A03333 | - |
dc.relation.journalcode | J02195 | - |
dc.identifier.eissn | 1741-0444 | - |
dc.identifier.pmid | 20440570 | - |
dc.identifier.url | http://link.springer.com/article/10.1007%2Fs11517-010-0621-2 | - |
dc.subject.keyword | Adjacent segment degeneration | - |
dc.subject.keyword | Pedicle screws | - |
dc.subject.keyword | Posterolateral fusion | - |
dc.subject.keyword | Finite element method | - |
dc.contributor.alternativeName | Kim, Hak Sun | - |
dc.contributor.alternativeName | Kim, Ho Joong | - |
dc.contributor.alternativeName | Moon, Seong Hwan | - |
dc.contributor.alternativeName | Moon, Eun Su | - |
dc.contributor.alternativeName | Park, Jin Oh | - |
dc.contributor.alternativeName | Lee, Hwan Mo | - |
dc.contributor.affiliatedAuthor | Kim, Ho Joong | - |
dc.contributor.affiliatedAuthor | Kim, Hak Sun | - |
dc.contributor.affiliatedAuthor | Moon, Seong Hwan | - |
dc.contributor.affiliatedAuthor | Moon, Eun Su | - |
dc.contributor.affiliatedAuthor | Park, Jin Oh | - |
dc.contributor.affiliatedAuthor | Lee, Hwan Mo | - |
dc.citation.volume | 48 | - |
dc.citation.number | 7 | - |
dc.citation.startPage | 703 | - |
dc.citation.endPage | 709 | - |
dc.identifier.bibliographicCitation | MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, Vol.48(7) : 703-709, 2010 | - |
dc.identifier.rimsid | 52684 | - |
dc.type.rims | ART | - |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.