optical quality measurements ; double-pass system ; ray tracing–type aberrometer ; visual acuity ; contrast sensitivity
Abstract
PURPOSE:
To compare the optical quality measurements obtained from the double-pass system and ocular aberrations, subjective visual acuity, and contrast sensitivity score in pseudophakic eyes.
METHODS:
Three months after cataract surgery, modulation transfer function (MTF) cutoff frequency, Strehl ratio, objective scatter index, and objective pseudoaccommodation obtained from the double-pass system were compared with total aberration, higher-order aberration, and spherical aberration obtained from ray-tracing aberrometer. In addition, parameters of the double-pass system were compared with subjective visual acuity and the contrast sensitivity score.
RESULTS:
Forty eyes of 40 patients were included. The MTF cutoff frequency and Strehl ratio were negatively correlated with total aberration (r = -0.503, p = 0.003; r = -0.509, p = 0.003, respectively) and subjective visual acuity (r = -0.453, p = 0.007; r = -0.354, p = 0.040, respectively). The objective scatter index was positively correlated with total aberration (r = 0.451, p = 0.024) and subjective visual acuity (r = 0.516, p = 0.008). The MTF cutoff frequency showed a correlation with contrast sensitivity score under photopic and mesopic conditions.
CONCLUSIONS:
Optical quality parameters obtained from the double-pass system were correlated with ocular aberrations, subjective visual acuity, and contrast sensitivity score in pseudophakic eyes.