Evaluation of performance including influence by interfering substances of the Innovance D-dimer assay on the Sysmex coagulation analyzer.

Authors
 Seo-Jin Park  ;  Hyun-Sook Chi  ;  Soh Hyun Chun  ;  Seongsoo Jang  ;  Chan-Jeoung Park 
Citation
 ANNALS OF CLINICAL AND LABORATORY SCIENCE, Vol.41(1) : 20-24, 2011 
Journal Title
ANNALS OF CLINICAL AND LABORATORY SCIENCE
ISSN
 0091-7370 
Issue Date
2011
MeSH
Adolescent ; Adult ; Aged ; Aged, 80 and over ; Bilirubin/metabolism* ; Blood Coagulation Tests/instrumentation* ; Blotting, Western ; Child ; Child, Preschool ; Female ; Fibrin Fibrinogen Degradation Products/analysis* ; Hemoglobins/metabolism* ; Humans ; Immunoassay/methods* ; Infant ; Infant, Newborn ; Male ; Middle Aged ; Young Adult
Abstract
D-dimer is formed during activation of the coagulation system and is commonly assayed in order to diagnose disseminated intravascular coagulation, deep vein thrombosis, and pulmonary embolism. Enzyme-linked immunosorbent assay has been validated as the reference method, but it is a time-consuming procedure. The objective of this study was to evaluate a new immunoturbidimetric, particle-enhanced, Innovance(®) D-dimer immunoassay. A total of 129 plasma samples from apparently healthy individuals and 298 samples from patients were collected for linearity, precision, and correlation studies. Testing the precision of low- and high-controls yielded CV values of 2.08% and 1.76%, respectively. The central 95% non-parametric reference interval estimated from healthy controls was 0.093-0.68 mg/L Fibrinogen Equivalent Units (FEU; median, 0.26 mg/L FEU). Comparison analysis yielded acceptable correlation with the STA Liatest(®) D-dimer assay (R(2) = 0.9471). At a cut-off level of <0.5 mg/L FEU, the sensitivity and specificity indices of the Innovance D-dimer assay were 99.7% and 89.1%, respectively. Thus the Innovance D-dimer method showed acceptable precision and linearity, and the assay results showed acceptable correlation with the STA Liatest D-dimer method. The Innovance method was relatively unaffected by potential interfering substances such as bilirubin and hemoglobin. In conclusion, the Innovance D-dimer assay is suitable for monitoring D-dimer concentrations in various clinical conditions and should be useful in clinical laboratories.
Full Text
http://www.annclinlabsci.org/content/41/1/20.long
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Laboratory Medicine (진단검사의학교실) > 1. Journal Papers
Yonsei Authors
Park, Seo Jin(박서진)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/95171
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links