Variations in dose distribution and optical properties of Gafchromic(TM) EBT2 film according to scanning mode.

Authors
 Soah Park  ;  Sei-Kwon Kang  ;  Kwang-Ho Cheong  ;  Taejin Hwang  ;  Haeyoung Kim  ;  Taejin Han  ;  Me-Yeon Lee  ;  KyoungJu Kim  ;  Hoonsik Bae  ;  Hyeong Su Kim  ;  Jung Han Kim  ;  Seung Jae Oh  ;  Jin-Suck Suh 
Citation
 MEDICAL PHYSICS, Vol.39(5) : 2524-2535, 2012 
Journal Title
MEDICAL PHYSICS
ISSN
 0094-2405 
Issue Date
2012
MeSH
Film Dosimetry/instrumentation ; Film Dosimetry/methods* ; Light ; Optical Phenomena* ; Radiation Dosage*
Keywords
EBT2 film ; transmission ; reflection ; optical properties ; rotation ; room light
Abstract
PURPOSE: The authors aim was to investigate the effects of using transmission and reflection scanning modes, the film orientation during scanning, and ambient room light on a dosimetry system based on the Gafchromic(TM) EBT2 film model.

METHODS: For calibration, the films were cut to 3 × 3 cm(2) and irradiated from 20 to 700 cGy at the depth of maximum dose using 6 and 10 MV photon beams in a 10 × 10 cm(2) field size. Absolute dose calibration of the linear accelerator was done according to the TRS398 protocol. An FG65-G ionization chamber was used to monitor the dose while irradiating the films in solid water. The film pieces were scanned with an EPSON Expression 1680 Pro flatbed scanner in transmission and reflection modes. Authors investigated the effect of orientation on films and examined the optical properties of EBT2 film using an ellipsometer and an ultraviolet (UV)/visible spectrometer to explain the dosimetric dependence of the film on orientation during the scanning process. To investigate the effect of ambient room light, films were preirradiated in 6 and 10 MV photon beams with intensity-modulated radiotherapy (IMRT) quality assurance (QA) plans, and then exposed to room light, either directly for 2 days in a workroom or for 2 months in a film box. Gamma index pass criteria of (3%, 3 mm) were used.

RESULTS: The dose response curves based on net optical density (NOD) indicated that the reflection scanning mode can provide a better dose sensitivity than the transmission scanning mode, whereas the standard deviation of the dose is greater in reflection mode than in transmission mode. When the film was rotated 90° from the portrait orientation, the average dose of the EBT2 film decreased by 11.5-19.6% in transmission mode and by 1.5-2.3% in reflection mode. Using an ellipsometer, variation of the refractive index of EBT2 film-the birefringence property-was found to be the largest between 45° (1.72 and 1.71) and 135° (1.8 and 1.77) for 300 and 800 cGy. Absorption spectra of EBT2 films measured with spectrometer were the function of film orientation. The readings in reflection scanning mode were more stable against room light than those in transmission scanning mode, although dose readings increased in both modes after the films were exposed to room light.

CONCLUSIONS: The transmission scanning mode exhibited a strong dependence on film orientation during scanning and a change in optical density resulting from room light exposure, so a constant scanning orientation and minimal exposure to light can reduce uncertainty in the measured dose (23 ± 3%). The angular dependence was analyzed using Jones matrices and optical properties of EBT2 film were obtained using an ellipsometer and an UV/visible spectrometer. The reflection scanning mode has relatively good stability with respect to room light and film orientation on a scanner, although the large standard deviation of dose is a disadvantage in measurements of absolute dose. Reflection scanning mode can offer a potential advantage for film dosimetry in radiotherapy, although transmission scanning mode is still recommended for dosimetry as it provides better uncertainty results.
Full Text
http://online.medphys.org/resource/1/mphya6/v39/i5/p2524_s1
DOI
22559623
Appears in Collections:
1. College of Medicine (의과대학) > Research Institute (부설연구소) > 1. Journal Papers
1. College of Medicine (의과대학) > Dept. of Radiology (영상의학교실) > 1. Journal Papers
Yonsei Authors
Suh, Jin Suck(서진석) ORCID logo https://orcid.org/0000-0001-9455-9240
Oh, Seung Jae(오승재)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/91444
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links