Cited 0 times in

Comparison of matrix-assisted laser desorption ionization–time-of-flight mass spectrometry assay with conventional methods for detection of IMP-6, VIM-2, NDM-1, SIM-1, KPC-1, OXA-23, and OXA-51 carbapenemase-producing Acinetobacter spp., Pseudomonas aeruginosa, and Klebsiella pneumoniae

Title
Comparison of matrix-assisted laser desorption ionization–time-of-flight mass spectrometry assay with conventional methods for detection of IMP-6, VIM-2, NDM-1, SIM-1, KPC-1, OXA-23, and OXA-51 carbapenemase-producing Acinetobacter spp., Pseudomonas aeruginosa, and Klebsiella pneumoniae
Authors
Wonmok Lee;Hae-Sun Chung;Yunsop Chong;Kyungwon Lee;Seok Hoon Jeong;Dongeun Yong;Yangsoon Lee
Issue Date
2013
Journal Title
Diagnostic Microbiology and Infectious Disease
ISSN
0732-8893
Citation
Diagnostic Microbiology and Infectious Disease, Vol.77(3) : 227~230, 2013
Abstract
Matrix-assisted laser desorption ionization-time-of-flight mass spectrometry assay was able to detect carbapenemase producers, including SIM-1 or OXA-51, within 4 hours using 20 μL of 0.5 g/L ertapenem solution as a substrate. This assay is more rapid and accurate than the modified Hodge test and 3-dimensional extract bioassay. Hence, it can be used an alternative test to identify carbapenemase-mediated carbapenem resistance in Gram-negative bacteria.
URI
http://www.sciencedirect.com/science/article/pii/S0732889313003982

http://ir.ymlib.yonsei.ac.kr/handle/22282913/88292
DOI
10.1016/j.diagmicrobio.2013.07.005
Appears in Collections:
1. 연구논문 > 1. College of Medicine > Dept. of Laboratory Medicine
Yonsei Authors
사서에게 알리기
  feedback
Files in This Item:
Export
RIS (EndNote)
XLS (Excel)
XML

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse