43 75

Cited 0 times in

Evaluation of a Robotic Liquid Handler, Andrew+, for Sample Preparation for LC-MS/MS-Based Quantitation of Immunosuppressants in Whole Blood

DC Field Value Language
dc.contributor.author이상국-
dc.date.accessioned2024-01-03T01:21:47Z-
dc.date.available2024-01-03T01:21:47Z-
dc.date.issued2023-07-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/197534-
dc.description.abstractBackground: Liquid chromatography/tandem mass spectrometry (LC-MS/MS)-based assays are increasingly being used for therapeutic drug monitoring (TDM) of immunosuppressive drugs (ISDs). However, many laboratories that employ LC-MS/MS assays for TDM of ISDs use manual sample preparation steps. Methods: We implemented a semi-automated sample preparation method based on protein precipitation, using an Andrew+ liquid handler, integrated with an LC-MS/MS-based TDM assay for four ISDs (tacrolimus, sirolimus, everolimus, and cyclosporine A). The feasibility and analytical performance were analyzed and compared with those of the manual sample preparation method using 163 clinical samples. Results: The total execution time of Andrew+-based sample preparation method was 55 minutes (including off-line centrifugation) for 48 samples. However, the hands-on time was significantly reduced than that of the manual sample preparation method. The average repeatability and within-laboratory precision of the assay were acceptable (CV ≤10%), and the results of manual and Andrew+-based sample preparation methods were equivalent. Conclusions: The sample preparation method using the Andrew+ liquid handler was successfully integrated with LC-MS/MS-based TDM assays for ISDs. Sample preparation using Andrew+ was reproducible, standardized, and enabled excellent utilization of technical staff time. Although further optimization for full automation is still pursued, this study indicates the potential of integrating liquid handers with LC-MS/MS-based assays in the clinical setting.-
dc.description.statementOfResponsibilityopen-
dc.languageKorean-
dc.publisherKorean Society for Laboratory Medicine-
dc.relation.isPartOfLaboratory Medicine Online-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.titleEvaluation of a Robotic Liquid Handler, Andrew+, for Sample Preparation for LC-MS/MS-Based Quantitation of Immunosuppressants in Whole Blood-
dc.title.alternative액체크로마토그래피-질량분석법 기반 전혈 내 면억억제제 약물농도 정량검사에 접목한 Andrew+ 검체 로봇 전처리방법 평가-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Laboratory Medicine (진단검사의학교실)-
dc.contributor.googleauthorHyojin Chae-
dc.contributor.googleauthorHyejeong Kim-
dc.contributor.googleauthorJeong Eun Kim-
dc.contributor.googleauthorJeong Joong Lee-
dc.contributor.googleauthorSung-Eun Cho-
dc.contributor.googleauthorSollip Kim-
dc.contributor.googleauthorKyunghoon Lee-
dc.contributor.googleauthorSang-Guk Lee-
dc.contributor.googleauthorYong-Wha Lee-
dc.contributor.googleauthorHyung-Doo Park-
dc.contributor.googleauthorSeungman Park-
dc.contributor.googleauthorSang Hoon Song-
dc.contributor.googleauthorSail Chun-
dc.contributor.googleauthorJunghan Song-
dc.contributor.googleauthorClinical Mass Spectrometry Research Committee (CMSRC)-
dc.contributor.googleauthorKorean Society of Clinical Chemistry (KSCC)-
dc.identifier.doi10.47429/lmo.2023.13.3.199-
dc.contributor.localIdA02810-
dc.relation.journalcodeJ02151-
dc.identifier.eissn2093-6338-
dc.subject.keywordAutomated liquid handling-
dc.subject.keywordImmunosuppressive drug-
dc.subject.keywordLiquid chromatography/tandem mass spectrometry-
dc.subject.keywordSample preparation-
dc.contributor.alternativeNameLee, Sang Guk-
dc.contributor.affiliatedAuthor이상국-
dc.citation.volume13-
dc.citation.number3-
dc.citation.startPage199-
dc.citation.endPage204-
dc.identifier.bibliographicCitationLaboratory Medicine Online, Vol.13(3) : 199-204, 2023-07-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Laboratory Medicine (진단검사의학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.