102 176

Cited 0 times in

자기공명영상을 위한 새로운 조직 상응 모형 의 제작과 평가

Other Titles
 The Production and Evaluation of the Tissue-equivalent Phantom for the Magnetic Resonance Imaging 
Authors
 유영훈  ;  조재현  ;  이재면  ;  정은기  ;  서진석 
Citation
 Journal of the Korean Radiologist Society (대한방사선의학회지), Vol.30(6) : 1151-1155, 1994-06 
Journal Title
Journal of the Korean Radiologist Society(대한방사선의학회지)
ISSN
 0301-2867 
Issue Date
1994-06
Abstract
PURPOSE: For the production and evaluation of the tissue-equivalent phantom, MATERIALS AND METHODS : We usedagarous gel and oil as a basic component of the mixture and added Tween 80 for the stabilization of phantoms, Wedid the test for homogeneity and measured T1 and T2 relexation times of each phantom tube. RESULTS: T1 relaxationtime ranged from 642 to 2781 msec and T2 relaxation times from 42 to 157 msec. Each phantom was significantlydifferent in T1 relaxation time and T2 relaxation time (p< .0001). CONCLUSION: Tissue equivalent phantom mayprovide good information on the optimal sequence before MR imaging of patients and may be valuable if it is usedwith the patients' MR imaging.
Files in This Item:
T199400866.pdf Download
DOI
10.3348/jkrs.1994.30.6.1151
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Nuclear Medicine (핵의학교실) > 1. Journal Papers
Yonsei Authors
Ryu, Young Hoon(유영훈) ORCID logo https://orcid.org/0000-0002-9000-5563
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/195000
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links