Phantom-based correction for standardization of myocardial native T1 and extracellular volume fraction in healthy subjects at 3-Tesla cardiac magnetic resonance imaging

DC Field Value Language
dc.contributor.author서영주-
dc.contributor.author최병욱-
dc.date.accessioned2023-03-03T02:19:17Z-
dc.date.available2023-03-03T02:19:17Z-
dc.date.issued2022-12-
dc.identifier.issn0938-7994-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/192799-
dc.description.abstractObjectives: To investigate the effect of the phantom-based correction method for standardizing myocardial native T1 and extracellular volume fraction (ECV) in healthy subjects. Methods: Seventy-one healthy asymptomatic adult (≥ 20 years) volunteers of five different age groups (34 men and 37 women, 45.5 ± 15.5 years) were prospectively enrolled in three academic hospitals. Cardiac MRI including Modified Look - Locker Inversion recovery T1 mapping sequence was performed using a 3-Tesla system with a different type of scanner for each hospital. Native T1 and ECV were measured in the short-axis T1 map and analyzed for mean values of the 16 entire segments. The myocardial T1 value of each subject was corrected based on the site-specific equation derived from the T1 Mapping and ECV Standardization phantom. The global native T1 and ECV were compared between institutions before and after phantom-based correction, and the variation in native T1 and ECV among institutions was assessed using a coefficient of variation (CoV). Results: The global native T1 value significantly differed between the institutions (1198.7 ± 32.1 ms, institution A; 1217.7 ± 39.9 ms, institution B; 1232.7 ± 31.1 ms, institution C; p = 0.002), but the mean ECV did not (26.6-27.5%, p = 0.355). After phantom-based correction, the global native T1 and ECV were 1289.7 ± 32.4 ms and 25.0 ± 2.7%, respectively, and CoV for native T1 between the three institutions decreased from 3.0 to 2.5%. The corrected native T1 value did not significantly differ between institutions (1284.5 ± 31.5 ms, institution A; 1296.5 ± 39.1 ms, institution B; 1291.3 ± 29.3 ms, institution C; p = 0.440), and neither did the ECV (24.4-25.9%, p = 0.078). Conclusions: The phantom-based correction method can provide standardized reference T1 values in healthy subjects. Key points: • After phantom-based correction, the global native T1 of 16 entire myocardial segments on 3-T cardiac MRI is 1289.4 ± 32.4 ms, and the extracellular volume fraction was 25.0 ± 2.7% for healthy subjects. • After phantom - based correction was applied, the differences in the global native T1 among institutions became insignificant, and the CoV also decreased from 3.0 to 2.5%.-
dc.description.statementOfResponsibilityopen-
dc.languageEnglish-
dc.publisherSpringer International-
dc.relation.isPartOfEUROPEAN RADIOLOGY-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.subject.MESHAdult-
dc.subject.MESHContrast Media-
dc.subject.MESHFemale-
dc.subject.MESHHealthy Volunteers-
dc.subject.MESHHumans-
dc.subject.MESHMagnetic Resonance Imaging* / methods-
dc.subject.MESHMagnetic Resonance Imaging, Cine-
dc.subject.MESHMale-
dc.subject.MESHMiddle Aged-
dc.subject.MESHMyocardium* / pathology-
dc.subject.MESHPredictive Value of Tests-
dc.subject.MESHReference Standards-
dc.subject.MESHReproducibility of Results-
dc.titlePhantom-based correction for standardization of myocardial native T1 and extracellular volume fraction in healthy subjects at 3-Tesla cardiac magnetic resonance imaging-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Radiology (영상의학교실)-
dc.contributor.googleauthorYoung Joo Suh-
dc.contributor.googleauthorPan Ki Kim-
dc.contributor.googleauthorJinho Park-
dc.contributor.googleauthorEun-Ah Park-
dc.contributor.googleauthorJung Im Jung-
dc.contributor.googleauthorByoung Wook Choi-
dc.identifier.doi10.1007/s00330-022-08936-8-
dc.contributor.localIdA01892-
dc.contributor.localIdA04059-
dc.relation.journalcodeJ00851-
dc.identifier.eissn1432-1084-
dc.identifier.pmid35771246-
dc.subject.keywordHeart-
dc.subject.keywordMagnetic resonance imaging-
dc.subject.keywordPhantoms, imaging-
dc.subject.keywordStandardization-
dc.contributor.alternativeNameSuh, Young Joo-
dc.contributor.affiliatedAuthor서영주-
dc.contributor.affiliatedAuthor최병욱-
dc.citation.volume32-
dc.citation.number12-
dc.citation.startPage8122-
dc.citation.endPage8130-
dc.identifier.bibliographicCitationEUROPEAN RADIOLOGY, Vol.32(12) : 8122-8130, 2022-12-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiology (영상의학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.