1072 1473

Cited 0 times in

IOL Master 700과 500, 초음파의 안구 생체계측 및 백내장 술 후 굴절력 비교

Other Titles
 Comparison of Ocular Biometry and Refractive Outcomes Using IOL Master 700, IOL Master 500, and Ultrasound 
Authors
 유태근  ;  최문정  ;  이형근  ;  서경률  ;  김응권  ;  김태임 
Citation
 JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY, Vol.58(5) : 523-529, 2017 
Journal Title
JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY(대한안과학회지)
ISSN
 0378-6471 
Issue Date
2017
Keywords
A-scan ; Biometry ; Cataract ; Intraocular lens ; IOL Master 700
Abstract
Purpose: To compare the new swept-source optical coherence tomography based IOL Master 700 to both the partial coherence interferometry based IOL Master 500 and ultrasound A-scan in terms of the ocular biometry and the prediction of postoperative refractive outcomes.
Methods: A total 67 eyes of 55 patients who received cataract surgery were included in our study. The axial length, anterior chamber depth, and keratometry were measured using IOL Master 700, IOL Master 500, and A-scan. The predictive errors, which are the differences between predictive refraction and post-operative refraction 1 month after surgery, were also compared.
Results: Axial length measurements were not successful in 5 eyes measured using IOL Master 700 and in 12 eyes measured using IOL Master 500. The mean absolute postoperative refraction predictive errors were 0.63 ± 0.50 diopters, 0.66 ± 0.51 diopters, and 0.62 ± 0.51 diopters for IOL Master 700, IOL Master 500, and A-scan, respectively, and these values exhibited no statistically significant differences. The mean axial lengths were 24.25 ± 2.41 mm, 24.24 ± 2.40 mm, and 24.22 ± 2.39 mm; the mean anterior chamber depths were 3.09 ± 0.39 mm, 3.17 ± 0.39 mm, and 3.15 ± 0.46 mm; and the mean keratometry values were 44.12 ± 1.82 diopters, 44.57 ± 2.10 diopters, and 43.98 ± 1.84 diopters for the IOL Master 700, IOL Master 500, and A-scan groups, respectively. None of these parameters showed statistically significant differences between the three groups. Regarding pair-wise comparison, there were significant differences between the IOL Master 700 and the other devices.
Conclusions: The ocular biometric measurements measured using IOL Master 700, IOL Master 500, and A-scan showed no significant differences. However, IOL Master 700 demonstrated a superior ability to successfully take biometric measurements compared to IOL Master 500. Therefore, IOL Master 700 is capable of measuring ocular biometry for cataract surgery in clinical practice
Files in This Item:
T201705606.pdf Download
DOI
10.3341/jkos.2017.58.5.523
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Ophthalmology (안과학교실) > 1. Journal Papers
Yonsei Authors
Kim, Eung Kweon(김응권) ORCID logo https://orcid.org/0000-0002-1453-8042
Kim, Tae-Im(김태임) ORCID logo https://orcid.org/0000-0001-6414-3842
Seo, Kyoung Yul(서경률) ORCID logo https://orcid.org/0000-0002-9855-1980
Lee, Hyung Keun(이형근) ORCID logo https://orcid.org/0000-0002-1123-2136
URI