200 771

Cited 0 times in

신장 방사선 섭취량 결정을 위한 conjugate view 방법에 대한 평가

DC Field Value Language
dc.contributor.author김유선-
dc.contributor.author김희중-
dc.contributor.author박해정-
dc.contributor.author윤미진-
dc.contributor.author이종두-
dc.date.accessioned2017-05-04T07:38:47Z-
dc.date.available2017-05-04T07:38:47Z-
dc.date.issued2005-
dc.identifier.issn1225-6714-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/147585-
dc.description.abstractPurpose: In order to obtain better quantitation of kidney uptake, this study is to evaluate a conjugate view method (CVM) using a geometric mean attenuation correction for kidney uptake and to compare it to Gate's method. Materials & Methods: We used a Monte Carlo code, SIMIND and a Zubal phantom, to simulate kidney uptake. SIMIND was both simulated with or without scatter for the Zubal phantom. Also, a real phantom test was carried out using a dual-head gamma camera. The activity of 0.5 mCi was infused into two small cylinder phantoms of 5 cm diameter, and then, they were inserted into a cylinder phantom of 20 cm diameter. The results by the CVM method were compared with ideal data without both of attenuation and scatter and with Gate's method. The CVM was performed with or without scatter correction. The Gate's method was performed without scatter correction and it was evaluated with regards to $0.12cm^{-1}\;and\;0.15cm^{-1}$ attenuation coefficients. Data were analyzed with comparisons of mean counts in the legions of interest (ROI), profiles drawn over kidney images and linear regression. Correlation coefficients were calculated with ideal data, as well. Results: In the case of the computer simulation, mean counts measured from ideal data, the CVM and the Gate's method were (right $998{\pm}209$, left: $896{\pm}249$), (right: $911{\pm}207$, left: $815{\pm}265$), and (right: $1065{\pm}267$, left: $1546{\pm}267$), respectively. The ideal data showed good correlation with the CVM and the correlation coefficients of the CVM, Gate's method were (right: 0.91, left: 0.93) and (right: 0.85, left: 0.90), respectively. Conclusion: The conjugate view method using geometric mean attenuation correction resulted in better accuracy than the Gate's method. In conclusion, the conjugate view method independent of renal depths may provide more accurate kidney uptake.-
dc.description.statementOfResponsibilityopen-
dc.formatapplication/pdf-
dc.languageKorean-
dc.publisher대한핵의학회-
dc.relation.isPartOfKorean Journal of Nuclear Medicine (대한핵의학회지)-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.rightshttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.subject.MESHConjugate view method-
dc.subject.MESHgeometric mean attenuation correction-
dc.subject.MESHscatter correction-
dc.subject.MESHkidney uptake-
dc.subject.MESHMonte Carlo simulation-
dc.title신장 방사선 섭취량 결정을 위한 conjugate view 방법에 대한 평가-
dc.title.alternativeEvaluation of a Conjugate View Method for Determination of Kidney Uptake-
dc.typeArticle-
dc.publisher.locationKorea-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Surgery (외과학교실)-
dc.contributor.departmentDept. of Radiology (영상의학교실)-
dc.contributor.departmentDept. of Nuclear Medicine (핵의학교실)-
dc.contributor.departmentDept. of Nuclear Medicine (핵의학교실)-
dc.contributor.departmentDept. of Nuclear Medicine (핵의학교실)-
dc.contributor.googleauthor봉정균-
dc.contributor.googleauthor윤미진-
dc.contributor.googleauthor이종두-
dc.contributor.googleauthor김희중-
dc.contributor.googleauthor손혜경-
dc.contributor.googleauthor권윤영-
dc.contributor.googleauthor박해정-
dc.contributor.googleauthor김유선-
dc.contributor.googleauthor김윤선-
dc.contributor.localIdA00785-
dc.contributor.localIdA01224-
dc.contributor.localIdA01730-
dc.contributor.localIdA02550-
dc.contributor.localIdA03138-
dc.relation.journalcodeJ02071-
dc.relation.journalsince1987~2005-
dc.relation.journalafter2006~ Nuclear Medicine and Molecular Imaging (핵의학 분자영상)-
dc.subject.keywordConjugate view method-
dc.subject.keywordgeometric mean attenuation correction-
dc.subject.keywordscatter correction-
dc.subject.keywordkidney uptake-
dc.subject.keywordMonte Carlo simulation-
dc.contributor.alternativeNameKim, Yu Seun-
dc.contributor.alternativeNameKim, Hee Joung-
dc.contributor.alternativeNamePark, Hae Jeong-
dc.contributor.alternativeNameYun, Mi Jin-
dc.contributor.alternativeNameLee, Jong Doo-
dc.citation.volume39-
dc.citation.number3-
dc.citation.startPage191-
dc.citation.endPage199-
dc.identifier.bibliographicCitationKorean Journal of Nuclear Medicine (대한핵의학회지), Vol.39(3) : 191-199, 2005-
dc.date.modified2017-05-04-
dc.identifier.rimsid40357-
dc.type.rimsART-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Nuclear Medicine (핵의학교실) > 1. Journal Papers
1. College of Medicine (의과대학) > Dept. of Radiology (영상의학교실) > 1. Journal Papers
1. College of Medicine (의과대학) > Dept. of Surgery (외과학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.