0 712

Cited 3 times in

Biomechanical characteristics of self-ligating brackets in a vertically displaced canine model: a finite element analysis

Authors
 S.-J. Kim  ;  Y.-H. Kwon  ;  C.-J. Hwang 
Citation
 ORTHODONTICS & CRANIOFACIAL RESEARCH, Vol.19(2) : 102-113, 2016 
Journal Title
ORTHODONTICS & CRANIOFACIAL RESEARCH
ISSN
 1601-6335 
Issue Date
2016
MeSH
Dental Alloys ; Dental Stress Analysis ; Finite Element Analysis* ; Humans ; Orthodontic Appliance Design ; Orthodontic Brackets* ; Orthodontic Wires
Keywords
finite element analysis ; self-ligating brackets ; vertically displaced canine
Abstract
OBJECTIVES: The objective of this study was to compare the biomechanical characteristics between two types of self-ligating brackets and conventional metal brackets using finite element analysis of a vertically displaced canine model focusing on the desired force on the canine and undesirable forces on adjacent teeth.
MATERIALS AND METHODS: Three-dimensional finite element models of the maxillary dentition with 1-mm, 2-mm, and 3-mm vertically displaced canines were constructed. Two different self-ligating brackets (In-Ovation C and Smart clip) and a conventional metal bracket (Micro-arch) were modeled. After a 0.016-inch NiTi (0.40 mm, round) wire was engaged, the displacement of each tooth was calculated using x-, y-, and z-coordinates, and the tensile and compressive stresses were calculated.
RESULTS: The extrusion and maximal tensile stress of the canine differed little between the three brackets, but the intrusion and minimal compressive stress values of the adjacent teeth differed considerably and were highest in the Smart clip and least in the In-Ovation C. The extrusion and maximal tensile stress of the canine in the 3-mm displacement model was less than that in the 2-mm displacement model, and the intrusion and minimal compressive stress of the adjacent teeth increased with the degree of displacement.
CONCLUSIONS: Self-ligating brackets were not superior to conventional brackets in leveling a vertically displaced canine. A continuous arch wire may not be recommended for leveling of severely displaced canines whether using self-ligating or conventional brackets.
Full Text
http://onlinelibrary.wiley.com/doi/10.1111/ocr.12119/abstract
DOI
10.1111/ocr.12119
Appears in Collections:
2. College of Dentistry (치과대학) > Dept. of Orthodontics (교정과학교실) > 1. Journal Papers
2. College of Dentistry (치과대학) > Others (기타) > 1. Journal Papers
Yonsei Authors
Kim, Sung Jin(김성진) ORCID logo https://orcid.org/0000-0003-4115-0403
Hwang, Chung Ju(황충주) ORCID logo https://orcid.org/0000-0003-3024-4551
URI