0 461

Cited 1 times in

Improvement of multi-channel isolated bio-potential amplifiers using the multiplexing

Authors
 Sun K. Yoo  ;  Nam H. Kim  ;  Byung C. Chang 
Citation
 PHYSIOLOGICAL MEASUREMENT, Vol.23(2) : 469-476, 2002 
Journal Title
PHYSIOLOGICAL MEASUREMENT
ISSN
 0967-3334 
Issue Date
2002
MeSH
Brain/physiology ; Electric Impedance ; Electroencephalography/instrumentation* ; Electroencephalography/methods* ; Equipment Design ; Humans
Abstract
We propose a simple isolated link circuit for a multi-channel bio-potential amplifier, by adapting the multiplexing/demultiplexing technique that is generally used in digital circuit design. The reduced number of isolation amplifiers, due to the multiplexing technique, results in an increase in the amount of isolation, as well as a reduction in manufacturing cost. The simple incorporation of a dead-zone in multiplexed timing, and a pull-down resistor in the multiplexed signal line, enables a multiple isolated bio-signal transmission over the shared isolation amplifier, with an affordable inter-channel cross talk.
Full Text
http://iopscience.iop.org/0967-3334/23/2/320/
DOI
10.1088/0967-3334/23/2/320
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Medical Engineering (의학공학교실) > 1. Journal Papers
Yonsei Authors
Kim, Nam Hyun(김남현)
Yoo, Sun Kook(유선국) ORCID logo https://orcid.org/0000-0002-6032-4686
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/144087
사서에게 알리기