Comparison of 3 marking techniques in preoperative assessment of toric intraocular lenses using a wavefront aberrometer

Authors
 Young Jae Woo  ;  Hun Lee  ;  Hong Seok Kim  ;  Eung Kweon Kim  ;  Kyoung Yul Seo  ;  Tae-im Kim 
Citation
 JOURNAL OF CATARACT AND REFRACTIVE SURGERY, Vol.41(6) : 1232-1240, 2015 
Journal Title
JOURNAL OF CATARACT AND REFRACTIVE SURGERY
ISSN
 0886-3350 
Issue Date
2015
MeSH
Aberrometry* ; Adult ; Aged ; Aged, 80 and over ; Astigmatism/physiopathology ; Astigmatism/surgery* ; Cataract/therapy ; Cornea/anatomy & histology* ; Corneal Wavefront Aberration/diagnosis ; Female ; Humans ; Lens Implantation, Intraocular/methods* ; Lenses, Intraocular* ; Male ; Middle Aged ; Phacoemulsification/methods* ; Preoperative Care ; Prospective Studies
Abstract
PURPOSE: To assess the accuracy of 3 corneal horizontal meridian marking techniques using the iTrace Surgical Workstation in preoperative assessment for toric intraocular lens (IOL) implantation.

SETTING: Department of Ophthalmology, Yonsei University College of Medicine, Seoul, South Korea.

DESIGN: Prospective comparative observational study.

METHODS: Patients were allocated into 3 groups. Three techniques-marking with a nonpendular marker with a surgeon's direct visualization, slitlamp-assisted marking with a pendulum-attached marker, and slitlamp-assisted marking with a horizontal slit beam-were used to perform the horizontal meridian marking. The marks were immediately documented and aligned with the Zaldivar toric caliper transparent topographic map. Using a horizontal reference meridian determined by the toric caliper, the accuracy of corneal horizontal marking techniques was quantitatively evaluated by determining the rotational deviation and vertical misalignment.

RESULTS: For rotational deviation, the pendular marking showed the fewest errors to the horizontal reference meridian (mean error -0.66 degrees; mean absolute error [MAE] 3.77 degrees). There was a significant difference between the pendular marking and the surgeon's direct visual marking. For the MAE, the pendular marking showed less deviation than the surgeon's direct visual marking and the horizontal slit-beam marking. With vertical misalignment, there was no significant difference between the 3 different corneal horizontal meridian marking techniques.

CONCLUSION: The slitlamp-assisted technique with the pendulum-attached marker was more precise than marking with a nonpendular marker with a surgeon's direct visualization and marking with a horizontal slit beam in preoperative assessment for toric IOLs.
Full Text
http://www.sciencedirect.com/science/article/pii/S0886335015005982
DOI
10.1016/j.jcrs.2014.09.045
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Ophthalmology (안과학교실) > 1. Journal Papers
Yonsei Authors
Kim, Eung Kweon(김응권) ORCID logo https://orcid.org/0000-0002-1453-8042
Kim, Tae-Im(김태임) ORCID logo https://orcid.org/0000-0001-6414-3842
Kim, Hong Seok(김홍석)
Seo, Kyoung Yul(서경률) ORCID logo https://orcid.org/0000-0002-9855-1980
Lee, Hun(이훈)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/141047
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links